Reduction of Radiation Exposure From C-Arm Fluoroscopy During Orthopaedic Trauma Operations With Introduction of Real-Time Dosimetry
- PMID: 26360538
- DOI: 10.1097/BOT.0000000000000442
Reduction of Radiation Exposure From C-Arm Fluoroscopy During Orthopaedic Trauma Operations With Introduction of Real-Time Dosimetry
Abstract
Objectives: The use of fluoroscopy for indirect guidance in orthopaedic trauma surgery has increased. The purpose of this investigation was to assess how real-time visualization of radiation exposure impacts dose levels during orthopaedic trauma operations.
Design: Observational comparative study.
Setting: Level 1 trauma center orthopaedic trauma surgery operating room.
Patients/participants: The participants in this study were 83 patients with fractures of the ankle, tibia, femur, or acetabulum receiving definitive surgical fixation of their fracture; children under 18 years of age were excluded from the study. Fellowship trained orthopaedic trauma surgeons, resident orthopaedic surgeons, radiology technicians, and scrub nurses involved in the operations on included fracture patients were also participants.
Intervention: Real-time radiation exposure feedback from the Philips DoseAware device.
Main outcome measurements: Radiation exposure from fluoroscopy compared between phase 1, during which participants were blinded to exposure levels, and phase 2, during which participants were able to see exposure levels in real time.
Results: Overall mean radiation exposure was decreased by 60% in phase 2 compared with phase 1 (P = 0.023). Mean surgeon (MS; average of primary and assistant surgeon) and mean nonsurgeon personnel (average of x-ray technician, scrub nurse, and patient) radiation exposures were decreased from phase 1 to phase 2, by 58% and 80%, respectively (MS, P = 0.034; mean nonsurgeon personnel, P = 0.043). From phase 1 to phase 2, MS radiation for femoral shaft fractures decreased by 80% or 162.0 μSv (P = 0.02) and by 81% or 128.9 μSv (P = 0.014) for acetabular fractures.
Discussion: Our data demonstrate that real-time visualization of radiation exposure during orthopaedic trauma operations can decrease radiation exposure in the highest exposure cases. Further research is necessary to determine whether the reduction in radiation exposure is sustained over time and to understand how real-time radiation exposure data mitigates exposure.
Level of evidence: Therapeutic Level III. See Instructions for Authors for a complete description of levels of evidence.
Similar articles
-
Prospective Assessment of the Oncogenic Risk to Patients From Fluoroscopy During Trauma Surgery.J Orthop Trauma. 2016 Jul;30(7):e223-9. doi: 10.1097/BOT.0000000000000543. J Orthop Trauma. 2016. PMID: 26825492
-
[Evaluation of radiation exposure of personnel in an orthopaedic and trauma operation theatre using the new real-time dosimetry system "dose aware"].Z Orthop Unfall. 2014 Aug;152(4):381-8. doi: 10.1055/s-0034-1368603. Epub 2014 Aug 21. Z Orthop Unfall. 2014. PMID: 25144849 German.
-
Percutaneous screw fixation of acetabular fractures with 2D fluoroscopy-based computerized navigation.Arch Orthop Trauma Surg. 2010 Sep;130(9):1177-83. doi: 10.1007/s00402-010-1095-2. Epub 2010 Apr 3. Arch Orthop Trauma Surg. 2010. PMID: 20364263
-
Image guidance: fluoroscopic navigation.Clin Orthop Relat Res. 2004 Apr;(421):70-6. Clin Orthop Relat Res. 2004. PMID: 15123929 Review.
-
C-arm fluoroscopy in orthopaedic surgical practice.Eur J Orthop Surg Traumatol. 2018 Dec;28(8):1563-1568. doi: 10.1007/s00590-018-2234-7. Epub 2018 May 23. Eur J Orthop Surg Traumatol. 2018. PMID: 29796825 Review.
Cited by
-
Learning to detect anatomical landmarks of the pelvis in X-rays from arbitrary views.Int J Comput Assist Radiol Surg. 2019 Sep;14(9):1463-1473. doi: 10.1007/s11548-019-01975-5. Epub 2019 Apr 20. Int J Comput Assist Radiol Surg. 2019. PMID: 31006106 Free PMC article.
-
Similar function and complications for patients with short versus long hip nailing for unstable pertrochanteric fractures.SICOT J. 2018;4:23. doi: 10.1051/sicotj/2018023. Epub 2018 Jun 15. SICOT J. 2018. PMID: 29905526 Free PMC article.
-
Radiation Safety Knowledge and Practices: Is the Indian Orthopaedic Community Well-informed?Indian J Orthop. 2020 Jul 31;54(Suppl 1):158-164. doi: 10.1007/s43465-020-00212-5. eCollection 2020 Sep. Indian J Orthop. 2020. PMID: 32952924 Free PMC article.
-
Radiation exposure for the surgical team in a hybrid-operating room.J Robot Surg. 2019 Feb;13(1):91-98. doi: 10.1007/s11701-018-0821-6. Epub 2018 May 10. J Robot Surg. 2019. PMID: 29748746
-
The Effect of C-Arm Mobility and Field of Vision on Radiation Exposure in the Treatment of Proximal Femoral Fractures: A Randomized Clinical Trial.Biomed Res Int. 2018 Mar 27;2018:6768272. doi: 10.1155/2018/6768272. eCollection 2018. Biomed Res Int. 2018. PMID: 29780829 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials