. 2015 Nov;31(11):1377-84.
doi: 10.1016/j.dental.2015.08.151. Epub 2015 Sep 8.

Flexural behavior of PEEK materials for dental application

Affiliations

Flexural behavior of PEEK materials for dental application

Andreas Dominik Schwitalla et al. Dent Mater. 2015 Nov.

Abstract

Objectives: The high-performance thermoplastic polymer PEEK (poly-ether-ether-ketone) is used as alternative implant material to metals since 1998 in many medical fields due to its bone-like mechanical properties. These iso-elastic characteristics of PEEK lead to the assumption, that it could represent a viable alternative to conventional materials also in the field of dentistry. Therefore the mechanical properties of different PEEK-compounds should be eavuated vie the three-point bending test.

Methods: 150 specimens of 11 different PEEK-compounds were tested via a three-point bending test using a universal testing machine (Z010, Zwick GmbH & Co, Ulm, Germany) after dry storage on the one hand and after incubation at 37°C in Ringer solution (Fa. Braun Melsungen AG, Melsungen, Deutschland) for one day, 7 days, 28 days und 84 days on the other hand and the material parameters bending modulus and bending strength were evaluated.

Results: The results regarding the bending moduli ranged from 2.73±0.26GPa for an unfilled brand to 47.27±10.3GPa for a carbon fiber reinforced PEEK with unidirectional continuous carbon fibers. Accordingly the bending strengths of these two material types ranged from 170.37±19.31MPa to 1009.63±107.33MPa.

Significance: All tested specimens showed higher values than the prevailing minimum strength for plastic materials and their application in dentistry of 65MPa (DIN EN ISO 10477). This underlines the applicability of PEEK in dentistry and points out the possibility to offer patients metal free restorations, especially in the presence of allergies and/or bruxism.

Keywords: Bending modulus; Bending strength; CFR-PEEK; PEEK; Prosthetic dentistry; Three-point bending test.