Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2016 Jun;26(6):1921-8.
doi: 10.1007/s00330-015-3985-8. Epub 2015 Sep 15.

Accelerated magnetic resonance diffusion tensor imaging of the median nerve using simultaneous multi-slice echo planar imaging with blipped CAIPIRINHA

Affiliations
Free article

Accelerated magnetic resonance diffusion tensor imaging of the median nerve using simultaneous multi-slice echo planar imaging with blipped CAIPIRINHA

Lukas Filli et al. Eur Radiol. 2016 Jun.
Free article

Abstract

Purpose: To investigate the feasibility of MR diffusion tensor imaging (DTI) of the median nerve using simultaneous multi-slice echo planar imaging (EPI) with blipped CAIPIRINHA.

Materials and methods: After federal ethics board approval, MR imaging of the median nerves of eight healthy volunteers (mean age, 29.4 years; range, 25-32) was performed at 3 T using a 16-channel hand/wrist coil. An EPI sequence (b-value, 1,000 s/mm(2); 20 gradient directions) was acquired without acceleration as well as with twofold and threefold slice acceleration. Fractional anisotropy (FA), mean diffusivity (MD) and quality of nerve tractography (number of tracks, average track length, track homogeneity, anatomical accuracy) were compared between the acquisitions using multivariate ANOVA and the Kruskal-Wallis test.

Results: Acquisition time was 6:08 min for standard DTI, 3:38 min for twofold and 2:31 min for threefold acceleration. No differences were found regarding FA (standard DTI: 0.620 ± 0.058; twofold acceleration: 0.642 ± 0.058; threefold acceleration: 0.644 ± 0.061; p ≥ 0.217) and MD (standard DTI: 1.076 ± 0.080 mm(2)/s; twofold acceleration: 1.016 ± 0.123 mm(2)/s; threefold acceleration: 0.979 ± 0.153 mm(2)/s; p ≥ 0.074). Twofold acceleration yielded similar tractography quality compared to standard DTI (p > 0.05). With threefold acceleration, however, average track length and track homogeneity decreased (p = 0.004-0.021).

Conclusion: Accelerated DTI of the median nerve is feasible. Twofold acceleration yields similar results to standard DTI.

Key points: • Standard DTI of the median nerve is limited by its long acquisition time. • Simultaneous multi-slice acquisition is a new technique for accelerated DTI. • Accelerated DTI of the median nerve yields similar results to standard DTI.

Keywords: Diffusion tensor imaging; Diffusion tractography; Echo-planar imaging; Median nerve; Simultaneous multi-slice.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Magn Reson Med. 2014 Sep;72(3):770-8 - PubMed
    1. Eur J Radiol. 2013 Oct;82(10):e590-6 - PubMed
    1. Eur Radiol. 2008 Oct;18(10):2283-91 - PubMed
    1. Eur Radiol. 2014 Mar;24(3):756-61 - PubMed
    1. AJNR Am J Neuroradiol. 2005 Oct;26(9):2183-6 - PubMed

LinkOut - more resources