Liver Steatosis Assessed by Controlled Attenuation Parameter (CAP) Measured with the XL Probe of the FibroScan: A Pilot Study Assessing Diagnostic Accuracy
- PMID: 26386476
- DOI: 10.1016/j.ultrasmedbio.2015.08.008
Liver Steatosis Assessed by Controlled Attenuation Parameter (CAP) Measured with the XL Probe of the FibroScan: A Pilot Study Assessing Diagnostic Accuracy
Abstract
To assess liver steatosis, the controlled attenuation parameter (CAP; giving an estimate of ultrasound attenuation ∼3.5 MHz) is available with the M probe of the FibroScan. We report on the adaptation of the CAP for the FibroScan XL probe (center frequency 2.5 MHz) without modifying the range of values (100-400 dB/m). CAP validation was successfully performed on Field II simulations and on tissue-mimicking phantoms. In vivo performance was assessed in a cohort of 59 patients spanning the range of steatosis. In vivo reproducibility was good and similar with both probes. The area under receiver operative characteristic curve was equal to 0.83/0.84 and 0.92/0.91 for the M/XL probes to detect >2% and >16% liver fat, respectively, as assessed by magnetic resonance imaging. Patients can now be assessed simultaneously for steatosis and fibrosis using the FibroScan, regardless of their morphology.
Keywords: Controlled attenuation parameter (CAP); Elastography; FibroScan; Liver; Non-alcoholic fatty liver disease (NAFLD); Steato-hepatitis; Steatosis; Ultrasound attenuation; Vibration-controlled transient elastography (VCTE).
Copyright © 2016 World Federation for Ultrasound in Medicine & Biology. Published by Elsevier Inc. All rights reserved.