Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2015 Oct;42(10):5735-44.
doi: 10.1118/1.4929939.

Theoretical detection threshold of the proton-acoustic range verification technique

Affiliations

Theoretical detection threshold of the proton-acoustic range verification technique

Moiz Ahmad et al. Med Phys. 2015 Oct.

Abstract

Purpose: Range verification in proton therapy using the proton-acoustic signal induced in the Bragg peak was investigated for typical clinical scenarios. The signal generation and detection processes were simulated in order to determine the signal-to-noise limits.

Methods: An analytical model was used to calculate the dose distribution and local pressure rise (per proton) for beams of different energy (100 and 160 MeV) and spot widths (1, 5, and 10 mm) in a water phantom. In this method, the acoustic waves propagating from the Bragg peak were generated by the general 3D pressure wave equation implemented using a finite element method. Various beam pulse widths (0.1-10 μs) were simulated by convolving the acoustic waves with Gaussian kernels. A realistic PZT ultrasound transducer (5 cm diameter) was simulated with a Butterworth bandpass filter with consideration of random noise based on a model of thermal noise in the transducer. The signal-to-noise ratio on a per-proton basis was calculated, determining the minimum number of protons required to generate a detectable pulse. The maximum spatial resolution of the proton-acoustic imaging modality was also estimated from the signal spectrum.

Results: The calculated noise in the transducer was 12-28 mPa, depending on the transducer central frequency (70-380 kHz). The minimum number of protons detectable by the technique was on the order of 3-30 × 10(6) per pulse, with 30-800 mGy dose per pulse at the Bragg peak. Wider pulses produced signal with lower acoustic frequencies, with 10 μs pulses producing signals with frequency less than 100 kHz.

Conclusions: The proton-acoustic process was simulated using a realistic model and the minimal detection limit was established for proton-acoustic range validation. These limits correspond to a best case scenario with a single large detector with no losses and detector thermal noise as the sensitivity limiting factor. Our study indicated practical proton-acoustic range verification may be feasible with approximately 5 × 10(6) protons/pulse and beam current.

PubMed Disclaimer

Figures

FIG. 1.
FIG. 1.
Dose distributions for the 100 MeV (left) and 160 MeV (right) beams. The dose values are the average dose per proton. The blue circular marker indicates the position of the detector (4 cm distal to the Bragg peak in each case).
FIG. 2.
FIG. 2.
Simulated acoustic waveforms at a detector 4 cm from the Bragg peak of the proton beam. The beam had a 100-MeV energy and a 1-mm spot size. Left: waveform for a beam with a 0.1-μs pulse width (235 mGy/pulse). Right: beam with a 10-μs pulse width (647 mGy/pulse). “Actual pressure” refers to the waveform detected by an ideal noiseless detector with infinite bandwidth. “Detected pressure” refers to the waveform measured by the realistically modeled transducer.
FIG. 3.
FIG. 3.
Spectrum of acoustic waveforms at a detector 4 cm from the Bragg peak of the proton beam. The beam had a 100-MeV energy and a 1-mm spot size. Left: spectrum for a beam with a 0.1-μs pulse width. Right: spectrum for a beam with a 10-μs pulse width. The dotted line is the magnitude of the transfer function of the transducer. Actual pressure refers to the spectrum detected by an ideal noiseless detector with infinite bandwidth. Detected pressure refers to the spectrum measured by the realistically modeled transducer.
FIG. 4.
FIG. 4.
Schematic of the non-inverting amplifier. The transducer is modeled as an input voltage source Vs with an electrical impedance Zs. V0 is the amplified signal output.
FIG. 5.
FIG. 5.
Noise equivalent hydrophone amplifier circuit.

References

    1. Lomax A. J., “Intensity modulated proton therapy and its sensitivity to treatment uncertainties 1: The potential effects of calculational uncertainties,” Phys. Med. Biol. 53(4), 1027–1042 (2008).10.1088/0031-9155/53/4/014 - DOI - PubMed
    1. Parodi K., Enghardt W., and Haberer T., “In-beam PET measurements of β+ radioactivity induced by proton beams,” Phys. Med. Biol. 47(1), 21–36 (2002).10.1088/0031-9155/47/1/302 - DOI - PubMed
    1. Min C.-H., Kim C. H., Youn M.-Y., and Kim J.-W., “Prompt gamma measurements for locating the dose falloff region in the proton therapy,” Appl. Phys. Lett. 89(18), 183517 (2006).10.1063/1.2378561 - DOI
    1. Peterson S. W., Robertson D., and Polf J., “Optimizing a three-stage Compton camera for measuring prompt gamma rays emitted during proton radiotherapy,” Phys. Med. Biol. 55(22), 6841–6856 (2010).10.1088/0031-9155/55/22/015 - DOI - PMC - PubMed
    1. Sulak L. et al., “Experimental studies of the acoustic signature of proton beams traversing fluid media,” Nucl. Instrum. Methods 161(2), 203–217 (1979).10.1016/0029-554X(79)90386-0 - DOI

Publication types