Effect of tear osmolarity on repeatability of keratometry for cataract surgery planning
- PMID: 26432124
- DOI: 10.1016/j.jcrs.2015.01.016
Effect of tear osmolarity on repeatability of keratometry for cataract surgery planning
Abstract
Purpose: To evaluate the effects of tear osmolarity on the repeatability of keratometry (K) measurements in patients presenting for cataract surgery.
Setting: Three clinical practices.
Design: Observational prospective nonrandomized study.
Methods: Subjects were prospectively recruited based on tear osmolarity (Tearlab Osmolarity System); that is, osmolarity more than 316 mOsm/L in at least 1 eye (hyperosmolar) and osmolarity less than 308 mOsm/L in both eyes (normal). The baseline K value was measured, and a second measurement was taken on the same instrument (IOLMaster) within 3 weeks of the first. Variability in average K, calculated corneal astigmatism using vector analysis, and intraocular lens (IOL) sphere power calculations were compared between groups.
Results: The hyperosmolar group (50 subjects) had a statistically significantly higher variability in the average K reading (P = .05) than the normal group (25 subjects) and a statistically significantly higher percentage of eyes with a 1.0 diopter (D) or greater difference in the measured corneal astigmatism (P = .02). A statistically significantly higher percentage of eyes in the hyperosmolar group had an IOL power difference of more than 0.5 D (P = .02). No statistically significant differences were present when the subjects were grouped by self-reported dry eye.
Conclusions: Significantly more variability in average K and anterior corneal astigmatism was observed in the hyperosmolar group, with significant resultant differences in IOL power calculations. Variability was not significantly different when subjects were grouped by self-reported dry eye. Measurement of tear osmolarity at the time of cataract surgery planning can effectively identify patients with a higher likelihood of high unexpected refractive error resulting from inaccurate keratometry.
Financial disclosure: Drs. Epitropoulos, Matossian, Berdy, and Malhotra received compensation from Tearlab for participating in the study. No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.