. 2016 Jul;26(7):2291-6.
doi: 10.1007/s00330-015-4051-2. Epub 2015 Oct 7.

Is there a systematic bias of apparent diffusion coefficient (ADC) measurements of the breast if measured on different workstations? An inter- and intra-reader agreement study

Affiliations

Is there a systematic bias of apparent diffusion coefficient (ADC) measurements of the breast if measured on different workstations? An inter- and intra-reader agreement study

Paola Clauser et al. Eur Radiol. 2016 Jul.

Abstract

Objectives: To evaluate the influence of post-processing systems, intra- and inter-reader agreement on the variability of apparent diffusion coefficient (ADC) measurements in breast lesions.

Methods: Forty-one patients with 41 biopsy-proven breast lesions gave their informed consent and were included in this prospective IRB-approved study. Magnetic resonance imaging (MRI) examinations were performed at 1.5 T using an EPI-DWI sequence, with b-values of 0 and 1000 s/mm(2). Two radiologists (R1, R2) reviewed the images in separate sessions and measured the ADC for lesion, using MRI-workstation (S-WS), PACS-workstation (P-WS) and a commercial DICOM viewer (O-SW). Agreement was evaluated using the intraclass correlation coefficient (ICC), Bland-Altman plots and coefficient of variation (CV).

Results: Thirty-one malignant, two high-risk and eight benign mass-like lesions were analysed. Intra-reader agreement was almost perfect (ICC-R1 = 0.974; ICC-R2 = 0.990) while inter-reader agreement was substantial (ICC from 0.615 to 0.682). Bland-Altman plots revealed a significant bias in ADC values measured between O-SW and S-WS (P = 0.025), no further systematic differences were identified. CV varied from 6.8 % to 7.9 %.

Conclusion: Post-processing systems may have a significant, although minor, impact on ADC measurements in breast lesions. While intra-reader agreement is high, the main source of ADC variability seems to be caused by inter-reader variation.

Key points: • ADC provides quantitative information on breast lesions independent from the system used. • ADC measurement using different workstations and software systems is generally reliable. • Systematic, but minor, differences may occur between different post-processing systems. • Inter-reader agreement of ADC measurements exceeded intra-reader agreement.

Keywords: Apparent Diffusion Coefficient; Breast; Diffusion-weighted Imaging; Magnetic Resonance Imaging; Observer Variation.