Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2016 May;43(5):1230-8.
doi: 10.1002/jmri.25073. Epub 2015 Oct 13.

Evaluation of optimized breath-hold and free-breathing 3D ultrashort echo time contrast agent-free MRI of the human lung

Affiliations

Evaluation of optimized breath-hold and free-breathing 3D ultrashort echo time contrast agent-free MRI of the human lung

Neville D Gai et al. J Magn Reson Imaging. 2016 May.

Abstract

Purpose: To evaluate an optimized stack of radials ultrashort echo time (UTE) 3D magnetic resonance imaging (MRI) sequence for breath-hold and free-breathing imaging of the human lung.

Materials and methods: A 3D stack of ultrashort echo time radials trajectory was optimized for coronal and axial lower-resolution breath-hold and higher-resolution free-breathing scans using Bloch simulations. The sequence was evaluated in 10 volunteers, without the use of contrast agents. Signal-to-noise ratio (SNR) mean and 95% confidence interval (CI) were determined from separate signal and noise images in a semiautomated fashion. The four scanning schemes were evaluated for significant differences in image quality using Student's t-test. Ten clinical patients were scanned with the sequence and findings were compared with concomitant computed tomography (CT) in nine patients. Breath-hold 3D spokes images were compared with 3D stack of radials in five volunteers. A Mann-Whitney U-test was performed to test significance in both cases.

Results: Breath-hold imaging of the entire lung in volunteers was performed with SNR (mean = 42.5 [CI]: 35.5-49.5; mean = 34.3 [CI]: 28.6-40) in lung parenchyma for coronal and axial scans, respectively, which can be used as a quick scout scan. Longer respiratory triggered free-breathing scan enabled high-resolution UTE scanning with mean SNR of 14.2 ([CI]: 12.9-15.5) and 9.2 ([CI]: 8.2-10.2) for coronal and axial scans, respectively. Axial free-breathing scans showed significantly higher image quality (P = 0.008) than the three other scanning schemes. The mean score for comparison with CT was 1.67 (score 0: n = 0; 1: n = 3; 2: n = 6). There was no significant difference between CT and MRI (P = 0.25). 3D stack of radials images were significantly better than 3D spokes images (P < 0.001).

Conclusion: The optimized 3D stack of radials trajectory was shown to provide high-quality MR images of the lung parenchyma without the use of MRI contrast agents. The sequence may offer the possibility of breath-hold imaging and provides greater flexibility in trading off slice thickness and parallel imaging for scan time.

Keywords: 3D stack of radials; breath-hold; free-breathing; lung parenchyma.

PubMed Disclaimer

Figures

Figure 1
Figure 1
(a): The simulated PSF in the x–y plane with the breath-hold scan parameters and typical lung relaxation parameters at 3T. Mxy = 1 corresponds to magnetization devoid of both T1 and T2 relaxation and for θ = 90°. The radius at FWHM was 0.85 mm (b) a 1-D profile of the PSF depicted at the center plane in Figure 1(a). FWHM was measured to be ~2 mm.
Figure 2
Figure 2
Single slice from coronal and axial lower resolution breath-hold (a) and higher resolution free breathing (b) scans of a normal volunteer.
Figure 3
Figure 3
MIP (25 mm slab thickness) of breath-hold lung images obtained at three different levels for (a) coronal and (b) axial scans. Native resolution was 2 × 2 × 8 mm3 for coronal scan and 2.5 × 2.5 × 8 mm3 for axial scan. Scan time was 18.5 s and 21 s, respectively, for complete coverage along the two directions.
Figure 4
Figure 4
MIP (25 mm slab thickness) of respiratory triggered free breathing lung images obtained at three different levels for (a) coronal and (b) axial scans. Native resolution was 1.5 × 1.5 × 5 mm3 for both scans. Scan time was 5:06 and 6:54 for complete coverage along the coronal and axial directions, respectively.
Figure 5
Figure 5
Semi-quantitative analysis of images for visualization of vessels and airways and the presence of artifacts. CL and AL refer to low resolution breath-hold coronal and axial images while CH and AH refer to high resolution FB coronal and axial scans, respectively. Note that maximum score for visualization was 4 (all the way to the periphery of lungs) while the best score for artifacts was 3 (no artifacts).
Figure 6
Figure 6
Scarring or atelectasis exhibited by a patient (CT and MRI done sequentially). (a) CT image compared with lower resolution breath-hold axial MRI image (red arrows) (b) CT image compared with free breathing axial MRI image (blue arrows). Although CT offers higher resolution, MR images provide comparable visualization.
Figure 7
Figure 7
Sample high resolution non-contrast MR image obtained with an isotropic resolution of 1 mm using a free breathing 3D STAR trajectory. Nominal scan time was 11 min 40 s. Sub-segmental airways are clearly visible on this image.
Figure 8
Figure 8
Comparison images obtained with the (a) 3D STAR and (b) the 3D spokes trajectory. Scan time was 20 s for each case. Resolution was fixed to the same voxel size (32 mm3) for each (2 × 2 × 8 mm3 for 3D STAR and 3.2 × 3.2 × 3.2 mm3 for 3D spokes.

References

    1. Hoffman EA, Lynch DA, Barr RG, van Beek EJ, Parraga G Investigators I. Pulmonary CT and MRI phenotypes that help explain chronic pulmonary obstruction disease pathophysiology and outcomes. J Magn Reson Imaging. 2015 Epub ahead of print. - PMC - PubMed
    1. Kauczor HU, Ley-Zaporozhan J, Ley S. Imaging of pulmonary pathologies: focus on magnetic resonance imaging. Proc Am Thorac Soc. 2009;6(5):458–463. - PubMed
    1. Wielputz M, Kauczor HU. MRI of the lung: state of the art. Diagn Interv Radiol. 2012;18(4):344–353. - PubMed
    1. Zhang WJ, Hubbard Cristinacce PL, Bondesson E, et al. MR Quantitative Equilibrium Signal Mapping: A Reliable Alternative to CT in the Assessment of Emphysema in Patients with Chronic Obstructive Pulmonary Disease. Radiology. 2015;275(2):579–588. - PubMed
    1. Atkinson IC, Lu A, Thulborn KR. Characterization and correction of system delays and eddy currents for MR imaging with ultrashort echo-time and time-varying gradients. Magn Reson Med. 2009;62(2):532–537. - PubMed

Publication types

MeSH terms