. 2016 Apr;21(2):287-96.
doi: 10.1177/2211068215610061. Epub 2015 Oct 13.

Automated Imaging and Analysis of the Hemagglutination Inhibition Assay

Affiliations

Automated Imaging and Analysis of the Hemagglutination Inhibition Assay

Michael Nguyen et al. J Lab Autom. 2016 Apr.

Abstract

The hemagglutination inhibition (HAI) assay quantifies the level of strain-specific influenza virus antibody present in serum and is the standard by which influenza vaccine immunogenicity is measured. The HAI assay endpoint requires real-time monitoring of rapidly evolving red blood cell (RBC) patterns for signs of agglutination at a rate of potentially thousands of patterns per day to meet the throughput needs for clinical testing. This analysis is typically performed manually through visual inspection by highly trained individuals. However, concordant HAI results across different labs are challenging to demonstrate due to analyst bias and variability in analysis methods. To address these issues, we have developed a bench-top, standalone, high-throughput imaging solution that automatically determines the agglutination states of up to 9600 HAI assay wells per hour and assigns HAI titers to 400 samples in a single unattended 30-min run. Images of the tilted plates are acquired as a function of time and analyzed using algorithms that were developed through comprehensive examination of manual classifications. Concordance testing of the imaging system with eight different influenza antigens demonstrates 100% agreement between automated and manual titer determination with a percent difference of ≤3.4% for all cases.

Keywords: automation; hemagglutination inhibition assay; high-throughput; image analysis; influenza.