The effect of implant thread design on stress distribution in anisotropic bone with different osseointegration conditions: a finite element analysis
- PMID: 26478976
- DOI: 10.11607/jomi.4091
The effect of implant thread design on stress distribution in anisotropic bone with different osseointegration conditions: a finite element analysis
Abstract
Purpose: Whereas bone is anisotropic, nearly all previous mechanical analyses of implants assumed bone as an isotropic material. Another means to simplify a simulation of the biomechanics of the implant-bone interface is the assumption of complete or no osseointegration; in clinical reality, an implant never achieves 100% contact with the surrounding bone. This study evaluated different thread profiles while not taking into account these two common simplifications. This study sought to (1) investigate the effects of various implant thread designs on stress distribution in the peri-implant bone, (2) appraise previous efforts in this area, and (3) find an optimum basic thread-form design.
Materials and methods: Through finite element analysis, four different basic commercial thread-form configurations for a solid screw-type implant were modeled: buttress, reverse buttress, V, and square. Bone was assumed to be transversely isotropic, and various degrees of osseointegration were simulated.
Results: Simulations showed that von Mises stresses were more distributed in the mesiodistal direction. Additionally, maximum stresses were concentrated at the cervical cortical bone region and the first thread. Moreover, in most of the models, von Mises stresses gradually increased in the supporting structure when the degree of osseointegration increased.
Conclusion: The use of different thread designs and various osseointegration conditions did not affect the stress distribution patterns in the supporting bone. In this study, square threads showed the most favorable results according to the predicted values of von Mises equivalent stress, pressure, different shear stresses, and micromotion.
Similar articles
-
The effect of thread design on stress distribution in a solid screw implant: a 3D finite element analysis.Clin Oral Investig. 2010 Aug;14(4):411-6. doi: 10.1007/s00784-009-0305-1. Epub 2009 Jun 20. Clin Oral Investig. 2010. PMID: 19543925
-
[Influence of thread shapes of custommade root-analogue implants on stress distribution of peri-implant bone: A three-dimensional finite element analysis].Beijing Da Xue Xue Bao Yi Xue Ban. 2019 Dec 18;51(6):1130-1137. doi: 10.19723/j.issn.1671-167X.2019.06.027. Beijing Da Xue Xue Bao Yi Xue Ban. 2019. PMID: 31848517 Free PMC article. Chinese.
-
Study and characterization of the crest module design: A 3D finite element analysis.J Prosthet Dent. 2015 Jun;113(6):541-7. doi: 10.1016/j.prosdent.2014.12.008. Epub 2015 Mar 18. J Prosthet Dent. 2015. PMID: 25794909
-
A review of improved fixation methods for dental implants. Part II: biomechanical integrity at bone-implant interface.J Prosthodont Res. 2015 Apr;59(2):84-95. doi: 10.1016/j.jpor.2015.01.003. Epub 2015 Mar 19. J Prosthodont Res. 2015. PMID: 25797023 Review.
-
Effect of different impactor designs on biomechanical behavior in the interface bone-implant: A comparative biomechanics study.Comput Methods Programs Biomed. 2020 Dec;197:105723. doi: 10.1016/j.cmpb.2020.105723. Epub 2020 Aug 25. Comput Methods Programs Biomed. 2020. PMID: 32877819 Review.
Cited by
-
Effect of short implant crown-to-implant ratio on stress distribution in anisotropic bone with different osseointegration rates.BMC Oral Health. 2023 Sep 20;23(1):683. doi: 10.1186/s12903-023-03379-z. BMC Oral Health. 2023. PMID: 37730562 Free PMC article.
-
Comparison of Implant Surgery Methods of Cortical Tapping and Cortical Widening in Bone of Various Density: A Three-Dimensional Finite Element Study.Materials (Basel). 2023 Apr 21;16(8):3261. doi: 10.3390/ma16083261. Materials (Basel). 2023. PMID: 37110097 Free PMC article.
-
Influence of three different implant thread designs on stress distribution: A three-dimensional finite element analysis.J Indian Prosthodont Soc. 2016 Oct-Dec;16(4):359-365. doi: 10.4103/0972-4052.191283. J Indian Prosthodont Soc. 2016. PMID: 27746600 Free PMC article.
-
Investigating the impact of diameters and thread designs on the Biomechanics of short implants placed in D4 bone: a 3D finite element analysis.BMC Oral Health. 2023 Sep 22;23(1):686. doi: 10.1186/s12903-023-03370-8. BMC Oral Health. 2023. PMID: 37740190 Free PMC article.
-
Three-dimensional finite element analysis of stress distribution on short implants with different bone conditions and osseointegration rates.BMC Oral Health. 2023 Apr 15;23(1):220. doi: 10.1186/s12903-023-02945-9. BMC Oral Health. 2023. PMID: 37061667 Free PMC article.
MeSH terms