Abstract
The theoretical basis and experimental implementation of FCS and FPR measurements are now well established. Because of the requirements for system stability and long data acquisition times FCS is relatively rarely used. But FCS can provide unique information, especially about extents of aggregation or polymerization and therefore is a useful supplement to FPR for certain applications. FPR measurements are now carried out routinely in many laboratories in a variety of formats using different beam profiles, optical systems, and analytical schemes. A particular version may be better adapted to a specific application. The spot photobleaching approach, however, seems simplest and most versatile for cellular studies and is now most often used. Important experimental considerations in setting up a spot photobleaching instrument are discussed in detail in Chapter 10 by Wolf (this volume) and elsewhere (Petersen et al., 1986a). In interpreting FPR measurements it is also important to take into account the possibility of systematic errors from a number of sources. In Chapter 10 in this volume, Wolf discusses many factors that must be properly controlled in carrying out FPR measurements. Additional consideration of some of these points is presented by Petersen et al. (1986a). One potentially troublesome type of error arises from the possibility that chemical reactions initiated by the photobleaching pulse or during the measurement of recovery could significantly perturb the system. Evidence from a variety of sources [summarized, for example, in Petersen et al. (1986a)] indicates that photobleaching fluorophores can induce chemical cross-linking of cellular proteins under some conditions. But measurements in a number of different systems have demonstrated that, even if these types of reactions occur in FPR measurements, nevertheless they do not perturb the measured mobilities. If possible, however, this point should be checked for each new system because variations in structure or environmental conditions could enhance the chemical cross-linking reactions mediated by photogenerated free radicals. In practice, the principal difficulty in carrying out FPR measurements on cells is frequently the low intensity of the fluorescent signal which can be obtained from specifically labeled cell surface ligands or microinjected components. This low intensity results from the typically low capacity of an individual cell for the specifically labeled macromolecule. Even in the absence of systematic errors, low emission intensity will reduce the accuracy of measurements due to shot noise. This is an important practical limitation on measuring accuracy. Low measurement accuracy severely limits the extent to which the data can be interpreted mechanistically. Precision can be improved by averaging many recovery experiments.(ABSTRACT TRUNCATED AT 400 WORDS)
Similar articles
-
Measurement of lateral transport on cell surfaces.Prog Clin Biol Res. 1976;9:137-47. Prog Clin Biol Res. 1976. PMID: 1030797
-
Interferometric fringe fluorescence photobleaching recovery interrogates entire cell surfaces.Biophys J. 1998 Aug;75(2):1131-8. doi: 10.1016/S0006-3495(98)77602-7. Biophys J. 1998. PMID: 9675214 Free PMC article.
-
High probe intensity photobleaching measurement of lateral diffusion in cell membranes.J Fluoresc. 2005 Nov;15(6):873-82. doi: 10.1007/s10895-005-0012-5. J Fluoresc. 2005. PMID: 16315103
-
Quick tour of fluorescence correlation spectroscopy from its inception.J Biomed Opt. 2004 Sep-Oct;9(5):857-64. doi: 10.1117/1.1779234. J Biomed Opt. 2004. PMID: 15447006 Review.
-
Accessing molecular dynamics in cells by fluorescence correlation spectroscopy.Biol Chem. 2001 Mar;382(3):491-4. doi: 10.1515/BC.2001.061. Biol Chem. 2001. PMID: 11347899 Review.
Cited by
-
Visualization of CD2 interaction with LFA-3 and determination of the two-dimensional dissociation constant for adhesion receptors in a contact area.J Cell Biol. 1996 Feb;132(3):465-74. doi: 10.1083/jcb.132.3.465. J Cell Biol. 1996. PMID: 8636222 Free PMC article.
-
Fluorescence correlation spectroscopy simulations of photophysical phenomena and molecular interactions: a molecular dynamics/monte carlo approach.J Phys Chem B. 2006 Feb 2;110(4):1896-906. doi: 10.1021/jp055840k. J Phys Chem B. 2006. PMID: 16471761 Free PMC article.
-
Theory of fluorescence recovery after photobleaching measurements on cylindrical surfaces.Biophys J. 1992 Feb;61(2):487-93. doi: 10.1016/S0006-3495(92)81853-2. Biophys J. 1992. PMID: 1547334 Free PMC article.
-
Neural activity controls the synaptic accumulation of alpha-synuclein.J Neurosci. 2005 Nov 23;25(47):10913-21. doi: 10.1523/JNEUROSCI.2922-05.2005. J Neurosci. 2005. PMID: 16306404 Free PMC article.
-
A supramolecular basis for CD45 tyrosine phosphatase regulation in sustained T cell activation.Proc Natl Acad Sci U S A. 2000 Aug 29;97(18):10138-43. doi: 10.1073/pnas.97.18.10138. Proc Natl Acad Sci U S A. 2000. PMID: 10963676 Free PMC article.