Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2016 Mar;30(3):254-6.
doi: 10.1089/end.2015.0642. Epub 2015 Dec 14.

Cost Analysis of Flexible Ureteroscope Repairs: Evaluation of 655 Procedures in a Community-Based Practice

Affiliations

Cost Analysis of Flexible Ureteroscope Repairs: Evaluation of 655 Procedures in a Community-Based Practice

Eugene Kramolowsky et al. J Endourol. 2016 Mar.

Abstract

Introduction: The frequency of flexible ureteroscopy has increased with the introduction of improved instrumentation. Ureteroscopes allow increased endoscopic access to the ureter and kidney. However, maintenance and repair of scopes may increase the total procedure expense.

Methods: In 3 years (8/2011-7/2014), 655 flexible ureteroscopies were performed at a single-specialty, urology, ambulatory surgery center. Procedures were performed by 26 board-certified urologists using four Olympus URF P5 flexible ureteroscopes. The instruments were handled by a single team and sterilized through the STERIS System E1. Repairs were performed by the manufacturer on an as needed basis. Patient records were reviewed to determine the preoperative diagnosis, operative time, location and size of the stone, and use of laser or ureteral sheath. The occurrence, nature of flexible ureteroscope damage, and cost of repairs were evaluated.

Results: Of the ureteroscopies performed, 78% was for the treatment of calculi (50.1% in the kidney). Mean stone size was 8.5 ± 0.2 mm, with larger stones (11 mm) located in the kidney. The flexible ureteroscope was advanced over a guidewire (88% of cases); a laser fiber was introduced in 70%, and a ureteral sheath was used in 13.4%. Mean procedure time was 40 minutes. The most common reasons for ureteroscope repair were cloudy lens (16 repairs) and broken optic fibers (9 repairs). There were 31 repairs during the study period (average 21 cases per repair). Flexible ureteroscopes were out of service for an average of 11 days per repair (range 3-20). The total cost of repairs was $233,150 or ∼$7521 per repair. The average repair cost per flexible ureteroscopy performed was $355.

Conclusions: Expenses associated with instrument repair can significantly impact a procedure's net revenue, thus efforts should be made to minimize instrument breakage. The expense of repairing a flexible ureteroscope per procedure can be significant and needs to be considered when pricing this procedure.