Comparison of macular choroidal thicknesses from swept source and spectral domain optical coherence tomography
- PMID: 26553923
- DOI: 10.1136/bjophthalmol-2015-307541
Comparison of macular choroidal thicknesses from swept source and spectral domain optical coherence tomography
Abstract
Background: Choroidal thickness (CT) measurements differ between swept source optical coherence tomography (SS-OCT) and spectral domain OCT (SD-OCT) devices for point thickness measurements. We aimed to assess the comparability of mean macular CT measurements between SS-OCT and SD-OCT devices.
Methods: In a prospective cohort study of 25 healthy volunteers, OCT scans were performed sequentially with the deep range imaging (DRI) OCT-1 and Spectralis OCT using standardised imaging protocols. These OCT scans were independently graded by reading centre-certified graders to obtain mean CT in the various Early Treatment Diabetic Retinopathy Study (ETDRS) subfields. Paired t tests and intraclass correlation coefficients (ICCs) were used to compare the measurements.
Results: The difference in mean central subfield CT between DRI OCT-1 and Spectralis was 49.3 µm (p<0.001), while differences in CT in various ETDRS subfields varied from 42.1 to 67.2 µm. After manual adjustment of the segmentation boundaries for the central subfield in the DRI OCT-1, the mean central subfield CT for DRI OCT-1 increased from 263.1 to 293.3 µm (p<0.001), and the resultant difference between DRI OCT-1 and Spectralis decreased from 49.3 to 19.1 µm (a decrease of 61.3%; p<0.001). CT between the three-dimensional and radial scanning protocols of the DRI OCT-1 were highly comparable, with differences generally under 10 µm and ICC of 0.888 for the central subfield.
Conclusions: CT measurements between automated segmentations from the DRI OCT-1 and manual segmentations on the Spectralis OCT may differ by more than 50 µm. This difference can be reduced, but not eliminated, by manual adjustment of segmentation boundaries by trained graders, and should be accounted for when comparing results between the two devices.
Keywords: Choroid; Diagnostic tests/Investigation; Imaging; Retina.
Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/.
Similar articles
-
Comparison of choroidal thicknesses using swept source and spectral domain optical coherence tomography in diseased and normal eyes.Br J Ophthalmol. 2015 Mar;99(3):354-8. doi: 10.1136/bjophthalmol-2014-305331. Epub 2014 Oct 1. Br J Ophthalmol. 2015. PMID: 25273828
-
Comparison of Macular Thickness Measurements Using Swept-Source and Spectral-Domain Optical Coherence Tomography in Healthy and Diabetic Subjects.Curr Eye Res. 2021 Oct;46(10):1567-1573. doi: 10.1080/02713683.2021.1908566. Epub 2021 Apr 20. Curr Eye Res. 2021. PMID: 33879001
-
A novel and faster method of manual grading to measure choroidal thickness using optical coherence tomography.Eye (Lond). 2018 Feb;32(2):433-438. doi: 10.1038/eye.2017.210. Epub 2017 Oct 20. Eye (Lond). 2018. PMID: 29052608 Free PMC article.
-
Comparison of retinal thicknesses measured using swept-source and spectral-domain optical coherence tomography devices.Ophthalmic Surg Lasers Imaging Retina. 2015 Feb;46(2):172-9. doi: 10.3928/23258160-20150213-23. Ophthalmic Surg Lasers Imaging Retina. 2015. PMID: 25707041
-
Swept-source optical coherence tomography imaging of the retinochoroid and beyond.Expert Rev Med Devices. 2020 May;17(5):413-426. doi: 10.1080/17434440.2020.1755256. Epub 2020 Apr 20. Expert Rev Med Devices. 2020. PMID: 32275451 Review.
Cited by
-
Optical coherence tomography evaluation of choroidal structure changes in diabetic retinopathy patients: A systematic review and meta-analysis.Front Med (Lausanne). 2022 Oct 20;9:986209. doi: 10.3389/fmed.2022.986209. eCollection 2022. Front Med (Lausanne). 2022. PMID: 36341274 Free PMC article.
-
Comparison of choroidal thickness measurements using swept source and spectral domain optical coherence tomography in pachychoroid diseases.PLoS One. 2020 Feb 26;15(2):e0229134. doi: 10.1371/journal.pone.0229134. eCollection 2020. PLoS One. 2020. PMID: 32101541 Free PMC article.
-
Choroidal thickness measured with swept source optical coherence tomography in posterior staphyloma strongly correlates with axial length and visual acuity.Int J Retina Vitreous. 2019 Jul 9;5:14. doi: 10.1186/s40942-019-0166-y. eCollection 2019. Int J Retina Vitreous. 2019. PMID: 31333879 Free PMC article.
-
Open-source algorithm for automatic choroid segmentation of OCT volume reconstructions.Sci Rep. 2017 Feb 9;7:42112. doi: 10.1038/srep42112. Sci Rep. 2017. PMID: 28181546 Free PMC article.
-
Predictors of persistent disease activity following anti-VEGF loading dose for nAMD patients in Singapore: the DIALS study.BMC Ophthalmol. 2020 Aug 6;20(1):324. doi: 10.1186/s12886-020-01582-y. BMC Ophthalmol. 2020. PMID: 32762659 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical