Auditing of Monitoring and Respiratory Support Equipment in a Level III-C Neonatal Intensive Care Unit
- PMID: 26558277
- PMCID: PMC4628988
- DOI: 10.1155/2015/719497
Auditing of Monitoring and Respiratory Support Equipment in a Level III-C Neonatal Intensive Care Unit
Abstract
Background: Random safety audits (RSAs) are a safety tool but have not been widely used in hospitals.
Objectives: To determine the frequency of proper use of equipment safety mechanisms in relation to monitoring and mechanical ventilation by performing RSAs. The study also determined whether factors related to the patient, time period, or characteristics of the area of admission influenced how the device safety systems were used.
Methods: A prospective observational study was conducted in a level III-C Neonatal Intensive Care Unit (NICU) during 2012. 87 days were randomly selected. Appropriate overall use was defined when all evaluated variables were correctly programmed in the audited device.
Results: A total of 383 monitor and ventilator audits were performed. The Kappa coefficient of interobserver agreement was 0.93. The rate of appropriate overall use of the monitors and respiratory support equipment was 33.68%. Significant differences were found with improved usage during weekends, OR 1.85 (1.12-3.06, p = 0.01), and during the late shift (3 pm to 10 pm), OR 1.59 (1.03-2.4, p = 0.03).
Conclusions: Equipment safety systems of monitors and ventilators are not properly used. To improve patient safety, we should identify which alarms are really needed and where the difficulties lie for the correct alarm programming.
Figures
Similar articles
-
[Real-time safety audits in a neonatal unit].An Pediatr (Barc). 2017 Sep;87(3):148-154. doi: 10.1016/j.anpedi.2016.08.005. Epub 2016 Oct 17. An Pediatr (Barc). 2017. PMID: 27765565 Spanish.
-
Smart pumps and random safety audits in a Neonatal Intensive Care Unit: a new challenge for patient safety.BMC Pediatr. 2015 Dec 11;15:206. doi: 10.1186/s12887-015-0521-6. BMC Pediatr. 2015. PMID: 26654316 Free PMC article.
-
The heuristics of nurse responsiveness to critical patient monitor and ventilator alarms in a private room neonatal intensive care unit.PLoS One. 2017 Oct 5;12(10):e0184567. doi: 10.1371/journal.pone.0184567. eCollection 2017. PLoS One. 2017. PMID: 28981515 Free PMC article.
-
Understanding neonatal ventilation: strategies for decision making in the NICU.Neonatal Netw. 2013 Jul-Aug;32(4):246-61. doi: 10.1891/0730-0832.32.4.246. Neonatal Netw. 2013. PMID: 23835544 Review.
-
Ventilation monitors and alarms.S Afr Med J. 1985 Mar 16;67(11):410-3. S Afr Med J. 1985. PMID: 3920763 Review.
References
-
- Yañez Gallardo R., Cuadra Olmos R. La técnica delphi y la investigación en los servicios de salud. Ciencia y Enfermeria. 2008;14(1):9–15.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources