Evaluation of Frequency-Doubling Technology Perimetry as a Means of Screening for Glaucoma and Other Eye Diseases Using the National Health and Nutrition Examination Survey
- PMID: 26562502
- DOI: 10.1001/jamaophthalmol.2015.4459
Evaluation of Frequency-Doubling Technology Perimetry as a Means of Screening for Glaucoma and Other Eye Diseases Using the National Health and Nutrition Examination Survey
Abstract
Importance: Glaucoma is a significant cause of global blindness and there are, as yet, no effective means of screening.
Objective: To assess the potential role of frequency-doubling technology (FDT) perimetry in screening for glaucoma using data collected as part of the National Health and Nutrition Examination Survey (NHANES).
Design, setting, and participants: Reanalysis of cross-sectional data of 6797 participants in the 2005-2008 cycles of the NHANES, which evaluated a sample of the noninstitutionalized US population with at least light-perception vision. A subset of optic nerve photographs were regraded by 3 glaucoma specialists in December 2012. Each participant underwent visual field testing, including FDT perimetry screening, and had fundus photographs taken.
Main outcomes and measures: Sensitivity and specificity of FDT perimetry to detect glaucoma, macular disease, or decreased visual acuity.
Results: A total of 5746 NHANES participants had optic images originally graded. We regraded 1201 images of 1073 eyes of 548 participants with initial cup-disc ratio (CDR) of 0.6 or greater and 423 images of 360 eyes of 180 randomly selected participants with initial CDR less than 0.6. Diagnoses of glaucoma by disc photograph were 1.6% (3 of 180) in the CDR less than 0.6 group and 31.4% (172 of 548) in the CDR of 0.6 or greater group. The sensitivity of FDT was 33% (95% CI, 0%-87%) and specificity was 77% (95% CI, 71%-84%). For the group with at least 1 CDR of 0.6 or greater, sensitivity of FDT was 66% (95% CI, 59%-73%) and specificity was 70% (95% CI, 66%-75%). When analyzed to give FDT credit for identifying glaucoma, macular disease, or decreased visual acuity, the sensitivity of the test was 80% (95% CI, 77%-83%) and the specificity was 83% (95% CI, 82%-84%). Approximately 25% of the NHANES population was not able to successfully complete FDT testing, representing screening failures and decreasing specificity.
Conclusions and relevance: Using the 2005-2008 waves of the NHANES as a model of population-based screening for eye disease, FDT perimetry lacks both sensitivity and specificity as a means of screening for glaucoma, the presence of retinal disease, or decreased acuity in a population-based setting. Given that no single test of glaucoma has yet been shown to be appropriate in a screening setting, to our knowledge, investigators should consider novel methods of detecting glaucoma or combinations of tests that might work better in a screening setting.
Comment in
-
Use of Multiple Tests Improves Screening for Glaucoma.JAMA Ophthalmol. 2016 Aug 1;134(8):947-8. doi: 10.1001/jamaophthalmol.2016.1541. JAMA Ophthalmol. 2016. PMID: 27309498 No abstract available.
-
Use of Multiple Tests Improves Screening for Glaucoma-Reply.JAMA Ophthalmol. 2016 Aug 1;134(8):948. doi: 10.1001/jamaophthalmol.2016.1552. JAMA Ophthalmol. 2016. PMID: 27309720 No abstract available.
Similar articles
-
Predictive value of frequency doubling technology perimetry for detecting glaucoma in a developing country.J Glaucoma. 2005 Apr;14(2):128-34. doi: 10.1097/01.ijg.0000151883.07232.54. J Glaucoma. 2005. PMID: 15741814
-
Performance of frequency-doubling technology perimetry in a population-based prevalence survey of glaucoma: the Tajimi study.Ophthalmology. 2007 Jan;114(1):27-32. doi: 10.1016/j.ophtha.2006.06.041. Epub 2006 Oct 27. Ophthalmology. 2007. PMID: 17070580
-
Glaucoma diagnostics.Acta Ophthalmol. 2013 Feb;91 Thesis 1:1-32. doi: 10.1111/aos.12072. Acta Ophthalmol. 2013. PMID: 23384049
-
Frequency doubling technology perimetry for the detection of glaucomatous visual field loss.Am J Ophthalmol. 2000 Mar;129(3):376-8. doi: 10.1016/s0002-9394(00)00352-4. Am J Ophthalmol. 2000. PMID: 10755955 Review. No abstract available.
-
Frequency-doubling technology perimetry.Ophthalmol Clin North Am. 2003 Jun;16(2):213-25. doi: 10.1016/s0896-1549(03)00011-7. Ophthalmol Clin North Am. 2003. PMID: 12809159 Review.
Cited by
-
Comparison of Glaucoma Detection Performance of Binocular Perimetry Screening Program Using imo Perimetry With Frequency Doubling Technology.Transl Vis Sci Technol. 2024 Aug 1;13(8):9. doi: 10.1167/tvst.13.8.9. Transl Vis Sci Technol. 2024. PMID: 39102239 Free PMC article.
-
FDT Perimetry for Glaucoma Detection in Comprehensive Health Checkup Service.J Ophthalmol. 2020 Mar 29;2020:4687398. doi: 10.1155/2020/4687398. eCollection 2020. J Ophthalmol. 2020. PMID: 32318280 Free PMC article.
-
Role of advanced technology in the detection of sight-threatening eye disease in a UK community setting.BMJ Open Ophthalmol. 2019 Dec 4;4(1):e000347. doi: 10.1136/bmjophth-2019-000347. eCollection 2019. BMJ Open Ophthalmol. 2019. PMID: 31909190 Free PMC article.
-
Diabetes, Triglyceride Levels, and Other Risk Factors for Glaucoma in the National Health and Nutrition Examination Survey 2005-2008.Invest Ophthalmol Vis Sci. 2016 Apr 1;57(4):2152-7. doi: 10.1167/iovs.15-18373. Invest Ophthalmol Vis Sci. 2016. PMID: 27111561 Free PMC article.
-
The prevalence and determinants of glaucoma among 40 years and older Saudi residents in the Riyadh Governorate (except the Capital) - A community based survey.Saudi J Ophthalmol. 2019 Oct-Dec;33(4):332-337. doi: 10.1016/j.sjopt.2019.02.006. Epub 2019 May 4. Saudi J Ophthalmol. 2019. PMID: 31920442 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical