. 2016 Dec;88(7):1077-1082.
doi: 10.1002/ccd.26340. Epub 2015 Nov 28.

Co-registration of angiography and intravascular ultrasound images through image-based device tracking

Affiliations

Co-registration of angiography and intravascular ultrasound images through image-based device tracking

Megha Prasad et al. Catheter Cardiovasc Interv. 2016 Dec.

Abstract

Objectives: To determine the feasibility of automated co-registration of angiography and intravascular ultrasound (IVUS) to facilitate integration of these two imaging modalities in a synchronous manner.

Background: IVUS provides cross-sectional imaging of coronary arteries but lacks overview of the vascular territory provided by angiography. Co-registration of angiography and IVUS would increase utility of IVUS in the clinical setting.

Methods: Forty-nine consecutive patients undergoing surveillance for cardiac allograft vasculopathy with angiography and IVUS of the left anterior descending artery (LAD) were enrolled. A pre-IVUS angiogram of the LAD was performed followed by an ECG-triggered fluoroscopy (ECGTF) during IVUS pullback at 0.5 mm/s using an automatic pullback device. ECGTF was used to track the IVUS catheter during pullback and establish a spatial relationship to the pre-IVUS angiogram. Angio-IVUS co-registration was performed with a research prototype (Siemens Healthcare, Germany) and accuracy was evaluated by distance mismatch between angiography and IVUS images at vessel bifurcations.

Results: Median age was 54 (44.5, 67) years. The population was 82.6% male with minimal risk factors. The median (IQR) co-registration distance mismatch measured at 108 bifurcations in 42 (85%) patients was 0.35 (0.00-1.16) mm. Seven patients were excluded due to inappropriate data acquisition (n = 3) and failure of tracking (n = 4), e.g., due to overlapping sternal wires. Estimated effective radiation dose for ECGTF was 0.09 mSv.

Conclusion: This study demonstrates the feasibility of angio-IVUS co-registration which may be used as a clinical tool for localizing IVUS cross-sections along an angiographic roadmap. © 2015 Wiley Periodicals, Inc.

Keywords: coronary artery disease; imaging; integration; novel.