Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 1977 May-Jun;5(3):128-36.
doi: 10.1097/00003246-197705000-00002.

Optimal flow pattern for mechanical ventilation of the lungs. Evaluation with a model lung

Comparative Study

Optimal flow pattern for mechanical ventilation of the lungs. Evaluation with a model lung

J F Dammann et al. Crit Care Med. 1977 May-Jun.

Abstract

The major components entering into the inspiratory pattern of various respirators were tested on a model lung which had an abnormally high airway resistance on one side. The tests consisted of simultaneous nitrogen washout curves from each lung separately utilizing two mass spectrometers. The components tested included a constant versus accelerating wave form and the presence, duration or absence of an end-inspiratory pause. Respirators tested included the Bennett MA1. Engström 300 and the Elema Schonander Servo tventilator 900. The results demonstrated the importance of an end inspiratory pause in improving gas distribution and efficiency of washout. No difference was found between a constant or accelerating air flow. Preliminary results in man appear to confirm the importance of an end inspiratory pause.