Observational Study
. 2016 Sep;100(9):1201-5.
doi: 10.1136/bjophthalmol-2015-307779. Epub 2015 Dec 16.

Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500

Affiliations
Observational Study

Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500

Ahmet Akman et al. Br J Ophthalmol. 2016 Sep.

Abstract

Purpose: To compare the measurements and failure rates obtained with a new swept source optical coherence tomography (OCT)-based biometry to IOLMaster 500.

Setting: Eye Clinic, Baskent University Faculty of Medicine, Ankara, Turkey.

Design: Observational cross-sectional study and evaluation of a new diagnostic technology.

Methods: 188 eyes of 101 subjects were included in the study. Measurements of axial length (AL), anterior chamber depth (ACD), corneal power (K1 and K2) and the measurement failure rate with the new Zeiss IOLMaster 700 were compared with those obtained with the IOLMaster 500. The results were evaluated using Bland-Altman analyses. The differences between both methods were assessed using the paired samples t test, and their correlation was evaluated by intraclass correlation coefficient (ICC).

Results: The mean age was 68.32±12.71 years and the male/female ratio was 29/72. The agreements between two devices were outstanding regarding AL (ICC=1.0), ACD (ICC=0.920), K1 (ICC=0.992) and K2 (ICC=0.989) values. IOLMaster 700 was able to measure ACD AL, K1 and K2 in all eyes within high-quality SD limits of the manufacturer. IOLMaster 500 was able to measure ACD in 175 eyes, whereas measurements were not possible in the remaining 13 eyes. AL measurements were not possible for 17 eyes with IOLMaster 500. Nine of these eyes had posterior subcapsular cataracts and eight had dense nuclear cataracts.

Conclusions: Although the agreement between the two devices was excellent, the IOLMaster 700 was more effective in obtaining biometric measurements in eyes with posterior subcapsular and dense nuclear cataracts.

Keywords: Diagnostic tests/Investigation; Optics and Refraction.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Bland–Altman plot for anterior chamber depth (ACD) comparing IOLMaster 700 with IOLMaster 500. The 95% limits of agreement were −0.33 to 0.17 mm. Dotted lines, ±1.96 SD.
Figure 2
Figure 2
Bland–Altman plot for axial length (AL) comparing IOLMaster 700 with IOLMaster 500. The 95% limits of agreement were −0.045 to 0.035 mm. Dotted lines, ±1.96 SD.
Figure 3
Figure 3
Bland–Altman plot for K1 value comparing IOLMaster 700 with IOLMaster 500. The 95% limits of agreement were −0.48 to 0.38 D. Dotted lines, ±1.96 SD.
Figure 4
Figure 4
Bland–Altman plot for K2 value comparing IOLMaster 700 with IOLMaster 500. The 95% limits of agreement were −0.61 to 0.45 D. Dotted lines, ±1.96 SD.

References

    1. Olsen T. Calculation of intraocular lens power: a review. Acta Ophthalmol Scand 2007;85:472–85. 10.1111/j.1600-0420.2007.00879.x - DOI - PubMed
    1. Rohrer K, Frueh BE, Wälti R, et al. . Comparison and evaluation of ocular biometry using a new noncontact optical low-coherence reflectometer. Ophthalmology 2009;116:2087–92. 10.1016/j.ophtha.2009.04.019 - DOI - PubMed
    1. Wang J-K, Chang S-W. Optical biometry intraocular lens power calculation using different formulas in patients with different axial lengths. Int J Ophthalmol 2013;6:150–4. 10.3980/j.issn.2222-3959.2013.02.08 - DOI - PMC - PubMed
    1. Vogel A, Dick HB, Krummenauer F. Reproducibility of optical biometry using partial coherence interferometry: intraobserver and interobserver reliability. J Cataract Refract Surg 2001;27:1961–8. 10.1016/S0886-3350(01)01214-7 - DOI - PubMed
    1. Fontes BM, Fontes BM, Castro E. Intraocular lens power calculation by measuring axial length with partial optical coherence and ultrasonic biometry. Arq Bras Oftalmol 2011;74:166–70. 10.1590/S0004-27492011000300004 - DOI - PubMed

Publication types