Screening for antibodies against Treponema pallidum with chemiluminescent microparticle immunoassay: analysis of discordant serology results and clinical characterization
- PMID: 26680646
- DOI: 10.1177/0004563215623806
Screening for antibodies against Treponema pallidum with chemiluminescent microparticle immunoassay: analysis of discordant serology results and clinical characterization
Abstract
Background: Traditionally, testing for syphilis has consisted of initial screening with a non-treponemal test, then retesting reactive specimens with a treponemal test. Recent availability of a chemiluminescent microparticle immunoassay for detecting antibodies against Treponema pallidum has led several laboratories in China to adopt chemiluminescent microparticle immunoassay for screening of syphilis, with subsequent testing of reactive serum samples with non-treponemal tests. We evaluated the utility of chemiluminescent microparticle immunoassay for routine screening of syphilis.
Methods: Antibodies against Treponema pallidum were screened in 20,550 serum samples using chemiluminescent microparticle immunoassay. Chemiluminescent microparticle immunoassay-positive samples were reflexively tested with rapid plasma reagin tests and Treponema pallidum particle agglutination assays. Dot-immunoblot assays were used to confirm results of chemiluminescent microparticle immunoassay-positive and Treponema pallidum particle agglutination-negative serum samples.
Results: Overall, 267 samples (1.3%) were chemiluminescent microparticle immunoassay-positive, and 185 (69.3%) of those chemiluminescent microparticle immunoassay-positive serum samples were also Treponema pallidum particle agglutination-positive. Samples' signal to cut-off ratio for chemiluminescent microparticle immunoassay correlated with diagnostic reliability, as greater samples' signal to cut-off ratio corresponded with greater concordance between chemiluminescent microparticle immunoassay and Treponema pallidum particle agglutination results. Dot-immunoblot testing of 82 chemiluminescent microparticle immunoassay-positive and Treponema pallidum particle agglutination-negative serum samples showed that 16 samples (19.5%) were Dot-immunoblot-positive, 28 (34.2%) were indeterminate and 38 (46.3%) were negative.
Conclusions: Because there is a certain percentage of false-positive results using chemiluminescent microparticle immunoassay for routine screening of syphilis, further analysis by Treponema pallidum particle agglutination is recommended to confirm diagnostic results. While in screening populations discrepancies between chemiluminescent microparticle immunoassay and Treponema pallidum particle agglutination results are quite prevalent, confirmation by immunoblot assay may be useful.
Keywords: Screening; Treponema pallidum; serological test; syphilis.
© The Author(s) 2016.
Similar articles
-
Consistency Between Treponema pallidum Particle Agglutination Assay and Architect Chemiluminescent Microparticle Immunoassay and Characterization of Inconsistent Samples.J Clin Lab Anal. 2015 Jul;29(4):281-4. doi: 10.1002/jcla.21765. Epub 2014 May 19. J Clin Lab Anal. 2015. PMID: 24840601 Free PMC article.
-
Comparison and evaluation of Abbott chemiluminescent microparticle immunoassay and ChIVD light-initiated chemiluminescent assay in the detection of Treponema pallidum antibody.J Clin Lab Anal. 2020 Jul;34(7):e23275. doi: 10.1002/jcla.23275. Epub 2020 Mar 4. J Clin Lab Anal. 2020. PMID: 32133697 Free PMC article.
-
Evaluation of the Lumipulse G TP-N Chemiluminescent Immunoassay as a Syphilis Screening Test.J Clin Microbiol. 2017 Nov;55(11):3236-3241. doi: 10.1128/JCM.00966-17. Epub 2017 Sep 6. J Clin Microbiol. 2017. PMID: 28878003 Free PMC article.
-
[Recent Advances in Laboratory Diagnosis of Syphilis].Mikrobiyol Bul. 2023 Jan;57(1):141-155. doi: 10.5578/mb.20239912. Mikrobiyol Bul. 2023. PMID: 36636853 Review. Turkish.
-
Sensitivity and Specificity of Treponemal-specific Tests for the Diagnosis of Syphilis.Clin Infect Dis. 2020 Jun 24;71(Suppl 1):S13-S20. doi: 10.1093/cid/ciaa349. Clin Infect Dis. 2020. PMID: 32578866 Free PMC article.
Cited by
-
Sexually Transmitted Infections Treatment Guidelines, 2021.MMWR Recomm Rep. 2021 Jul 23;70(4):1-187. doi: 10.15585/mmwr.rr7004a1. MMWR Recomm Rep. 2021. PMID: 34292926 Free PMC article.
-
Using Treponemal Assay Signal Strength Cutoff Ratios To Predict Syphilis Infection.J Clin Microbiol. 2017 Dec 26;56(1):e01555-17. doi: 10.1128/JCM.01555-17. Print 2018 Jan. J Clin Microbiol. 2017. PMID: 29046409 Free PMC article.
-
The Traditional or Reverse Algorithm for Diagnosis of Syphilis: Pros and Cons.Clin Infect Dis. 2020 Jun 24;71(Suppl 1):S43-S51. doi: 10.1093/cid/ciaa307. Clin Infect Dis. 2020. PMID: 32578864 Free PMC article.
-
A retrospective study to assess the accuracy of syphilis rapid test by chemiluminescent microparticle immunoassay (CIA) in volunteer blood donors.Indian J Sex Transm Dis AIDS. 2025 Jan-Jun;46(1):22-24. doi: 10.4103/ijstd.ijstd_115_24. Epub 2025 Jun 9. Indian J Sex Transm Dis AIDS. 2025. PMID: 40546366 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical