Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2016 May;43(5):824-831.
doi: 10.1007/s00259-015-3277-1. Epub 2015 Dec 19.

Occupational radiation exposure of medical staff performing ⁹⁰Y-loaded microsphere radioembolization

Affiliations

Occupational radiation exposure of medical staff performing ⁹⁰Y-loaded microsphere radioembolization

Sophie Laffont et al. Eur J Nucl Med Mol Imaging. 2016 May.

Abstract

Purpose: Radioembolization of liver cancer with (90)Y-loaded microspheres is increasingly used but data regarding hospital staff exposure are scarce. We evaluated the radiation exposure of medical staff while preparing and injecting (90)Y-loaded glass and resin microspheres especially in view of the increasing use of these products.

Methods: Exposure of the chest and finger of the radiopharmacist, nuclear medicine physician and interventional radiologist during preparation and injection of 78 glass microsphere preparations and 16 resin microsphere preparations was monitored. Electronic dosimeters were used to measure chest exposure and ring dosimeters were used to measure finger exposure.

Results: Chest exposure was very low for both products used (<10 μSv from preparation and injection). In our experience, finger exposure was significantly lower than the annual limit of 500 mSv for both products. With glass microspheres, the mean finger exposure was 13.7 ± 5.2 μSv/GBq for the radiopharmacist, and initially 17.9 ± 5.4 μSv/GBq for the nuclear medicine physician reducing to 13.97 ± 7.9 μSv/GBq with increasing experience. With resin microspheres, finger exposure was more significant: mean finger exposure for the radiopharmacist was 295.1 ± 271.9 μSv/GBq but with a reduction with increasing experience to 97.5 ± 35.2 μSv/GBq for the six most recent dose preparations. For administration of resin microspheres, the greatest mean finger exposure for the nuclear medicine physician (the most exposed operator) was 235.5 ± 156 μSv/GBq.

Conclusion: Medical staff performing (90)Y-loaded microsphere radioembolization procedures are exposed to safe levels of radiation. Exposure is lower than that from treatments using (131)I-lipiodol. The lowest finger exposure is from glass microspheres. With resin microspheres finger exposure is acceptable but could be optimized in accordance with the ALARA principle, and especially in view of the increasing use of radioembolization.

Keywords: 90Y; Exposure; Microspheres; Radioembolization.

PubMed Disclaimer

Comment in

References

    1. Hepatology. 2011 Sep 2;54(3):868-78 - PubMed
    1. Semin Nucl Med. 2008 Mar;38(2):S40-5 - PubMed
    1. Lancet Oncol. 2012 Oct;13(10):1025-34 - PubMed
    1. Eur J Nucl Med Mol Imaging. 2013 Jul;40(7):1057-68 - PubMed
    1. Nucl Med Commun. 2008 Apr;29(4):405-10 - PubMed

Publication types

MeSH terms

LinkOut - more resources