Repeatability and comparability of corneal power and corneal astigmatism obtained from a point-source color light-emitting diode topographer, a Placido-based corneal topographer, and a low-coherence reflectometer
- PMID: 26703301
- DOI: 10.1016/j.jcrs.2015.11.003
Repeatability and comparability of corneal power and corneal astigmatism obtained from a point-source color light-emitting diode topographer, a Placido-based corneal topographer, and a low-coherence reflectometer
Abstract
Purpose: To assess the repeatability and agreement of corneal power and astigmatism obtained from the Cassini point-source color light-emitting diode (LED) topographer, Humphrey Atlas 9000 Placido-based corneal topographer, and Lenstar LS-900 low-coherence reflectometer in normal eyes.
Setting: Cullen Eye Institute, Baylor College of Medicine, Houston, Texas, USA.
Design: Evaluation of diagnostic test or technology.
Methods: Consecutively enrolled patients with normal corneas were enrolled. Three sets of measurements were obtained using the color-LED topographer, the Placido topographer, and the reflectometer. Vector analysis was used in the astigmatism analysis. The repeatability was evaluated using the within-subject standard deviation, coefficient of variation (CoV), and intraclass correlation coefficient (ICC). Agreement was verified using Bland-Altman plots. The paired Student t test was used to assess statistical significance.
Results: Thirty-two eyes (32 patients) were evaluated. All devices provided highly repeatable corneal power and astigmatism measurements (ICC > 0.9) except for the Placido topographer with regard to J45 (ICC = 0.721). The color-LED topographer and the reflectometer obtained similar mean values of corneal power, astigmatism magnitude, J0, and J45 (P > .05), which was also true when comparing the color-LED topographer and the Placido topographer, except for the mean corneal power (P = .0007). The Bland-Altman plots showed a wide data spread for all analyzed variables.
Conclusions: The color-LED topographer provided highly repeatable corneal power and astigmatism measurements. Even though it obtained values similar to those of the reflectometer and the Placido topographer, the wide data spread discourages their interchangeable use to assess corneal power and astigmatism.
Financial disclosure: Drs. Wang, Koch, and Weikert are consultants to Ziemer USA, Inc. Dr. Koch is a consultant to Abbott Medical Optics, Inc., Alcon Surgical, Inc., and i-Optics, Corp. Drs. Ventura and Al-Mohtaseb have no financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Comparison of corneal power, astigmatism, and wavefront aberration measurements obtained by a point-source color light-emitting diode-based topographer, a Placido-disk topographer, and a combined Placido and dual Scheimpflug device.J Cataract Refract Surg. 2015 Aug;41(8):1658-71. doi: 10.1016/j.jcrs.2014.11.051. J Cataract Refract Surg. 2015. PMID: 26432123
-
Repeatability of posterior and total corneal curvature measurements with a dual Scheimpflug-Placido tomographer.J Cataract Refract Surg. 2015 Dec;41(12):2731-8. doi: 10.1016/j.jcrs.2015.07.035. J Cataract Refract Surg. 2015. PMID: 26796454
-
Comparison of Color Light-Emitting Diode Corneal Topographer and Dual Rotating Scheimpflug-Placido Topographer.J Ophthalmol. 2018 Dec 27;2018:6764805. doi: 10.1155/2018/6764805. eCollection 2018. J Ophthalmol. 2018. PMID: 30687546 Free PMC article.
-
Evaluation of optical quality of corneas using corneal topographers.Cornea. 2002 Oct;21(7 Suppl):S75-8. doi: 10.1097/01.ico.0000263123.53521.4d. Cornea. 2002. PMID: 12484703 Review.
-
Contribution of the ocular surface to visual optics.Exp Eye Res. 2004 Mar;78(3):417-25. doi: 10.1016/j.exer.2003.10.012. Exp Eye Res. 2004. PMID: 15106921 Review.
Cited by
-
Agreement between a new fully automatic ocular biometer based on optical low-coherence reflectometry and an optical biometer based on Scheimpflug imaging combined with partial coherence interferometry.BMC Ophthalmol. 2024 Oct 17;24(1):455. doi: 10.1186/s12886-024-03716-y. BMC Ophthalmol. 2024. PMID: 39420260 Free PMC article.
-
Repeatability and reproducibility of a new spectral-domain optical coherence tomography biometer and agreement with swept-source optical coherence tomography based biometer.Eye Vis (Lond). 2025 Feb 1;12(1):6. doi: 10.1186/s40662-024-00422-0. Eye Vis (Lond). 2025. PMID: 39891211 Free PMC article.
-
A novel color-LED corneal topographer to assess astigmatism in pseudophakic eyes.Clin Ophthalmol. 2016 Aug 13;10:1521-9. doi: 10.2147/OPTH.S113027. eCollection 2016. Clin Ophthalmol. 2016. PMID: 27574391 Free PMC article.
-
Repeatability and comparability of the Galilei-G4 and Cassini in measuring corneal power and astigmatism in normal and post-refractive surgery eyes.Sci Rep. 2021 Aug 9;11(1):16141. doi: 10.1038/s41598-021-94319-w. Sci Rep. 2021. PMID: 34373485 Free PMC article.
-
Current Developments in Corneal Topography and Tomography.Diagnostics (Basel). 2021 Aug 13;11(8):1466. doi: 10.3390/diagnostics11081466. Diagnostics (Basel). 2021. PMID: 34441401 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources