An 83-year-old man presented with a 1-year history of blurry vision in his left eye. Bilateral cataract surgery and ciliary sulcus posterior chamber intraocular lens (IOL) implantation had been performed 11 years earlier. The corrected distance visual acuity (CDVA) was 20/60 in the left eye. Slitlamp examination revealed moderate posterior capsule opacification. Neodymium:YAG laser posterior capsulotomy was performed and complicated by IOL fracture and immediate dislocation of optic fragments into the anterior chamber. An IOL exchange was performed, and the damaged IOL was removed in 7 pieces. A 3-piece acrylic IOL was suture fixated to the posterior iris. Pathologic analysis showed that the damaged IOL was made of glass supported by a polyimide frame. Twelve months after the IOL exchange, the CDVA was 20/60.
Financial disclosure: Dr. Ahmed is a consultant to Alcon Laboratories, Inc., Abbott Medical Optics, Inc., and Bausch & Lomb. Dr. Mamalis is a consultant to Anew Optics, Inc., and Medennium, Inc. No other author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Haptic fracture and dislocation of polyimide intraocular lens after neodymium: YAG laser capsulotomy: A case report.Medicine (Baltimore). 2019 May;98(20):e15720. doi: 10.1097/MD.0000000000015720. Medicine (Baltimore). 2019. PMID: 31096526 Free PMC article.
-
Economic considerations related to choice of intraocular lens (IOL) and posterior capsule opacification frequency - a comparison of three different IOLs.Acta Ophthalmol. 2014 Mar;92(2):179-83. doi: 10.1111/aos.12026. Epub 2012 Dec 31. Acta Ophthalmol. 2014. PMID: 23280286
-
Posterior capsular opacification and Nd:YAG capsulotomy rates with the iMics Y-60H and Micro AY intra-ocular lenses: 3-year results of a randomized clinical trial.Acta Ophthalmol. 2015 Jun;93(4):342-7. doi: 10.1111/aos.12543. Epub 2014 Nov 13. Acta Ophthalmol. 2015. PMID: 25393894 Clinical Trial.
-
Negative dysphotopsia: A perfect storm.J Cataract Refract Surg. 2015 Oct;41(10):2291-312. doi: 10.1016/j.jcrs.2015.09.002. J Cataract Refract Surg. 2015. PMID: 26703307 Review.
-
[Posterior YAG capsulotomy: selection of the application pattern].Ophthalmologe. 2022 May;119(5):481-490. doi: 10.1007/s00347-021-01526-x. Epub 2021 Nov 12. Ophthalmologe. 2022. PMID: 34767082 Review. German.
Cited by
-
Clinical Outcomes of Primary Posterior Continuous Curvilinear Capsulorhexis in Postvitrectomy Cataract Eyes.J Ophthalmol. 2020 Jul 30;2020:6287274. doi: 10.1155/2020/6287274. eCollection 2020. J Ophthalmol. 2020. PMID: 32802488 Free PMC article.
-
Anti-tumor Drug THZ1 Suppresses TGFβ2-mediated EMT in Lens Epithelial Cells via Notch and TGFβ/Smad Signaling Pathway.J Cancer. 2019 Jun 9;10(16):3778-3788. doi: 10.7150/jca.30359. eCollection 2019. J Cancer. 2019. PMID: 31333795 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources