Comparison of 3 biometry devices in cataract patients
- PMID: 26703487
- DOI: 10.1016/j.jcrs.2015.05.028
Comparison of 3 biometry devices in cataract patients
Abstract
Purpose: Accurate biometry is an obligatory preoperative measurement for refractive surgery as well as cataract surgery. A new device based on partial coherence interferometry was compared with 2 currently used biometry devices.
Setting: Department of Ophthalmology, Saarland University Medical Center, Homburg, Germany.
Design: Prospective case series.
Methods: Eyes of cataract patients were examined with a functional prototype of the new optical low-coherence reflectometry (OLCR) biometer OA-2000, the standard OLCR biometer Lenstar, and the partial coherence interferometry (PCI) biometer IOLMaster. The results were compared using a Wilcoxon-Mann-Whitney U test and Pearson correlation calculations.
Results: A total of 138 eyes of 74 cataract patients were examined. Pearson correlation showed excellent correlation for axial length, anterior chamber depth and keratometry among the 3 devices tested. The highest correlation was found between standard OLCR biometer and PCI biometer for AL, R1, and R2 (r = 1.0, r = 0.936, r = 0.952, respectively; all P ≤ .001). For anterior chamber depth (ACD), the highest correlation was found between the standard OLCR biometer and the new OLCR biometer (r = 0.943; P ≤ .001). The mean values of AL/ACD/R1/R2 differed very little, but the differences were significant (all P ≤ .05) (new OLCR biometer 23.31/3.21/7.74/7.64 mm; standard OLCR biometer 23.30/3.13/7.80/7.60 mm; PCI biometer 23.37/3.00/7.78/7.6 mm).
Conclusions: Compared with other clinical instruments, the new OLCR biometer generated the most accurate results. Differences especially in measurement of axial length were statistically but not clinically significant. The new OLCR biometer yielded results that correlated very well with the values of the PCI biometer and standard OLCR biometer.
Financial disclosure: None of the authors has a financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Measurement agreement between a new biometer based on partial coherence interferometry and a validated biometer based on optical low-coherence reflectometry.J Cataract Refract Surg. 2016 Jan;42(1):68-75. doi: 10.1016/j.jcrs.2015.05.042. J Cataract Refract Surg. 2016. PMID: 26948780
-
Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry.J Cataract Refract Surg. 2016 Jan;42(1):76-83. doi: 10.1016/j.jcrs.2015.07.039. J Cataract Refract Surg. 2016. PMID: 26948781
-
Biometry and intraocular lens power calculation results with a new optical biometry device: comparison with the gold standard.J Cataract Refract Surg. 2014 Apr;40(4):593-600. doi: 10.1016/j.jcrs.2013.09.015. J Cataract Refract Surg. 2014. PMID: 24680520
-
Technical failure rates for biometry between swept-source and older-generation optical coherence methods: a review and meta-analysis.BMC Ophthalmol. 2023 Apr 26;23(1):182. doi: 10.1186/s12886-023-02926-0. BMC Ophthalmol. 2023. PMID: 37101115 Free PMC article. Review.
-
Cataract surgery in the small eye.J Cataract Refract Surg. 2015 Nov;41(11):2565-75. doi: 10.1016/j.jcrs.2015.10.008. J Cataract Refract Surg. 2015. PMID: 26703508 Review.
Cited by
-
Repeatability of a new swept-source optical coherence tomographer and agreement with other three optical biometers.Graefes Arch Clin Exp Ophthalmol. 2022 Jul;260(7):2271-2281. doi: 10.1007/s00417-022-05579-9. Epub 2022 Feb 16. Graefes Arch Clin Exp Ophthalmol. 2022. PMID: 35171331
-
Repeatability and Agreement of Central Vault for Implantable Collamer Lens Obtained by the Tomey OA-2000 Biometer and Spectralis OCT.J Ophthalmol. 2024 Jun 24;2024:3684626. doi: 10.1155/2024/3684626. eCollection 2024. J Ophthalmol. 2024. PMID: 38957378 Free PMC article.
-
Comparison of OA-2000 and IOL Master 500 using in cataract patients with high myopia.Int J Ophthalmol. 2019 May 18;12(5):844-847. doi: 10.18240/ijo.2019.05.23. eCollection 2019. Int J Ophthalmol. 2019. PMID: 31131247 Free PMC article.
-
Accuracy of intraocular lens power calculation using three optical biometry measurement devices: the OA-2000, Lenstar-LS900 and IOLMaster-500.Eye (Lond). 2018 Jul;32(7):1244-1252. doi: 10.1038/s41433-018-0063-x. Epub 2018 Mar 12. Eye (Lond). 2018. PMID: 29527012 Free PMC article.
-
Comparison of a new optical biometer and a standard biometer in cataract patients.Eye Vis (Lond). 2016 Oct 17;3:27. doi: 10.1186/s40662-016-0059-1. eCollection 2016. Eye Vis (Lond). 2016. PMID: 27833928 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous