Quicker response results in better SpO2 control - a comparison of 3 FiO2-titration strategies in ventilated preterm infants
- PMID: 26706983
- DOI: 10.5604/12321966.1185781
Quicker response results in better SpO2 control - a comparison of 3 FiO2-titration strategies in ventilated preterm infants
Abstract
Introduction: The impact of SpO2 target ranges (TR) has been carefully studied; however, reports suggest a wide variation among infants and centres in maintaining the intended range. Little is known about the effectiveness of different approaches to manual control. Auto-SpO2 controllers are now available which show promise.
Objective: The aim was to compare two different protocol-driven manual strategies with different response requirements to each other, and a faster automated system (AveaCLiO2, Yorba Linda, CA, USA).
Materials and methods: In a crossover design, each of the three FiO2/SpO2 approaches was implemented in three randomly assigned consecutive 2.5-hour runs. The two manual strategies (Attentive and Observational) were implemented by a trained operator. The primary endpoints were time in 1) SpO2 TR, 2) < 80% SpO2 and 3) >98% SpO2.
Results: Fifteen studies were completed. All three approaches resulted in good control, with time in the target range >60%. CLiO2 use reflected reduced exposure at the two SpO2 extremes. Post hoc analysis determined that the differences were more marked in the infants with more frequent desaturations. Likewise, in this group, the Attentive strategy performed better than the Observative.
Conclusions: All three approaches provided excellent control of SpO2 in infants with infrequent desaturations, significantly better than typical routine care. In hard to manage infants with frequent desaturations, faster response appeared to result in better control. The potential of automating the tedious error prone FiO2 adjustment offers significant promise. If manual titration of FiO2 is to remain the usual method of care, additional studies are needed to identify optimal approaches.
Similar articles
-
Automated FiO2-SpO2 control system in neonates requiring respiratory support: a comparison of a standard to a narrow SpO2 control range.BMC Pediatr. 2014 May 28;14:130. doi: 10.1186/1471-2431-14-130. BMC Pediatr. 2014. PMID: 24885124 Free PMC article. Clinical Trial.
-
USING THE AUTOMATED FIO2- SPO2 CONTROL IN NEONATAL INTENSIVE CARE UNITS IN POLAND. A PRELIMINARY REPORT.Dev Period Med. 2015 Jul-Sep;19(3 Pt 1):263-70. Dev Period Med. 2015. PMID: 26958688
-
Evaluation of two SpO2 alarm strategies during automated FiO2 control in the NICU: a randomized crossover study.BMC Pediatr. 2019 May 6;19(1):142. doi: 10.1186/s12887-019-1496-5. BMC Pediatr. 2019. PMID: 31060536 Free PMC article. Clinical Trial.
-
Automated versus manual control of inspired oxygen to target oxygen saturation in preterm infants: a systematic review and meta-analysis.J Perinatol. 2018 Apr;38(4):351-360. doi: 10.1038/s41372-017-0037-z. Epub 2018 Jan 2. J Perinatol. 2018. PMID: 29296004
-
Automated control of inspired oxygen (FiO2 ) in preterm infants: Literature review.Pediatr Pulmonol. 2019 Mar;54(3):358-363. doi: 10.1002/ppul.24238. Epub 2019 Jan 10. Pediatr Pulmonol. 2019. PMID: 30632296 Review.
Cited by
-
Automated oxygen delivery for preterm infants with respiratory dysfunction.Cochrane Database Syst Rev. 2023 Nov 30;11(11):CD013294. doi: 10.1002/14651858.CD013294.pub2. Cochrane Database Syst Rev. 2023. PMID: 38032241 Free PMC article.
-
Reducing the time delay of oxygen transport to the neonate on continuous positive airway pressure support: A bench study.Front Pediatr. 2023 Apr 19;11:1141432. doi: 10.3389/fped.2023.1141432. eCollection 2023. Front Pediatr. 2023. PMID: 37152305 Free PMC article.
-
Automated FiO2-SpO2 control system in neonates requiring respiratory support: a comparison of a standard to a narrow SpO2 control range.BMC Pediatr. 2014 May 28;14:130. doi: 10.1186/1471-2431-14-130. BMC Pediatr. 2014. PMID: 24885124 Free PMC article. Clinical Trial.
-
The effectiveness of automated adjustment of inspired oxygen in preterm infants receiving respiratory support compared with manual: A systematic review and meta-analysis.Pediatr Discov. 2024 May 22;2(4):e57. doi: 10.1002/pdi3.57. eCollection 2024 Dec. Pediatr Discov. 2024. PMID: 40626118 Free PMC article. Review.