. 2016 Jul;74(4):317-25.
doi: 10.1016/j.pharma.2015.12.003. Epub 2016 Jan 13.

[Development and validation of an analytical method to quantify residues of cleaning products able to inactivate prion]

[Article in French]
Affiliations

[Development and validation of an analytical method to quantify residues of cleaning products able to inactivate prion]

[Article in French]
T Briot et al. Ann Pharm Fr. 2016 Jul.

Abstract

Objectives: In this study, a novel analytical method to quantify prion inactivating detergent in rinsing waters coming from the washer-disinfector of a hospital sterilization unit has been developed. The final aim was to obtain an easy and functional method in a routine hospital process which does not need the cleaning product manufacturer services.

Methods: An ICP-MS method based on the potassium dosage of the washer-disinfector's rinsing waters was developed. Potassium hydroxide is present on the composition of the three prion inactivating detergent currently on the French market. The detergent used in this study was the Actanios LDI(®) (Anios laboratories). A Passing and Bablok regression compares concentrations measured with this developed method and with the HPLC-UV manufacturer method.

Results: According to results obtained, the developed method is easy to use in a routine hospital process. The Passing and Bablok regression showed that there is no statistical difference between the two analytical methods during the second rinsing step. Besides, both methods were linear on the third rinsing step, with a 1.5ppm difference between the concentrations measured for each method.

Conclusions: This study shows that the ICP-MS method developed is nonspecific for the detergent, but specific for the potassium element which is present in all prion inactivating detergent currently on the French market. This method should be functional for all the prion inactivating detergent containing potassium, if the sensibility of the method is sufficient when the potassium concentration is very low in the prion inactivating detergent formulation.

Keywords: ICP-MS; Laveur désinfecteur; Potassium; Prion; Sterilization; Stérilisation; Washer-disinfector.

PubMed Disclaimer