Deformation pattern and load transfer of an uncemented femoral stem with modular necks. An experimental study in human cadaver femurs

Affiliations

Deformation pattern and load transfer of an uncemented femoral stem with modular necks. An experimental study in human cadaver femurs

Cathrine H Enoksen et al. Clin Biomech (Bristol). 2016 Feb.

Abstract

Background: Modular necks in hip arthroplasty allow variations in neck-shaft angles, neck version and neck lengths and have been introduced to improve accuracy when reconstructing the anatomy and hip joint biomechanics. Periprosthetic bone resorption may be a consequence of stress shielding in the proximal femur after implantation of a femoral stem. The purpose of this study was to investigate the deformation pattern and load transfer of an uncemented femoral stem coupled to different modular necks in human cadaver femurs.

Methods: A cementless femoral stem was implanted in twelve human cadaver femurs and tested in a hip simulator corresponding to single leg stance and stair climbing activity with patient-specific loading. The stems were tested with four different modular necks; long, short, retro and varus. The long neck was used as reference in statistical comparisons, as it can be considered the "standard" neck. The deformation of bone during loading was measured by strain gauge rosettes at three levels of the proximal femur on the medial, lateral and anterior side.

Findings: The cortical strains were overall reduced on the medial and lateral side of femur, for all implants tested, and in both loading conditions compared to the unoperated femur. Although there were statistical significant differences between the necks, the results did not show a consistent pattern considering which neck retained or lost most strain. In general the differences were small, with the highest significant difference between the necks of 3.23 percentage points.

Interpretation: The small differences of strain between the modular necks tested in this study are not expected to influence bone remodeling in the proximal femur.

Keywords: Cortical strain; Deformation pattern; Femoral offset; In vitro study; Load transfer; Modular necks; Neck version; Total hip arthroplasty.