Effects of Different Insertion Techniques of a Cochlear Implant Electrode on the Intracochlear Pressure
- PMID: 26794906
- DOI: 10.1159/000442041
Effects of Different Insertion Techniques of a Cochlear Implant Electrode on the Intracochlear Pressure
Abstract
To achieve a functional atraumatic insertion, low intracochlear pressure changes during the procedure are assumed to be important. The aim of this study was to observe intracochlear pressure changes due to different insertion techniques in a cochlear model. Cochlear implant electrode insertions were performed in an artificial cochlear model to record intracochlear pressure changes with a micropressure sensor to evaluate the maximum amplitude and frequency of pressure changes under different insertional conditions. We found statistically significant differences in the occurrence of intracochlear pressure peak changes comparing different techniques. Based on our model results, an insertion should be maximally supported to minimize micromovement-related pressure changes.
© 2016 S. Karger AG, Basel.
Similar articles
-
Postinsertional Cable Movements of Cochlear Implant Electrodes and Their Effects on Intracochlear Pressure.Biomed Res Int. 2016;2016:3937196. doi: 10.1155/2016/3937196. Epub 2016 Nov 9. Biomed Res Int. 2016. PMID: 27900329 Free PMC article.
-
Optimisation of the round window opening in cochlear implant surgery in wet and dry conditions: impact on intracochlear pressure changes.Eur Arch Otorhinolaryngol. 2016 Nov;273(11):3609-3613. doi: 10.1007/s00405-016-3993-6. Epub 2016 Mar 18. Eur Arch Otorhinolaryngol. 2016. PMID: 26993657
-
Cochlear implant electrode sealing techniques and related intracochlear pressure changes.J Otolaryngol Head Neck Surg. 2017 May 11;46(1):40. doi: 10.1186/s40463-017-0218-y. J Otolaryngol Head Neck Surg. 2017. PMID: 28490377 Free PMC article.
-
Cochlear implant electrode insertion: in defence of cochleostomy and factors against the round window membrane approach.Cochlear Implants Int. 2011 Aug;12 Suppl 2:S36-9. doi: 10.1179/146701011X13074645127478. Cochlear Implants Int. 2011. PMID: 21917217 Review.
-
Cochleostomy site: implications for electrode placement and hearing preservation.Acta Otolaryngol. 2005 Aug;125(8):870-6. doi: 10.1080/00016480510031489. Acta Otolaryngol. 2005. PMID: 16158535 Review.
Cited by
-
Intracochlear Pressure Transients During Cochlear Implant Electrode Insertion.Otol Neurotol. 2016 Dec;37(10):1541-1548. doi: 10.1097/MAO.0000000000001232. Otol Neurotol. 2016. PMID: 27753703 Free PMC article.
-
Combining Intraoperative Electrocochleography with Robotics-Assisted Electrode Array Insertion.Otol Neurotol. 2024 Feb 1;45(2):143-149. doi: 10.1097/MAO.0000000000004094. Otol Neurotol. 2024. PMID: 38206061 Free PMC article.
-
Effect of Underwater Insertion on Intracochlear Pressure.Front Surg. 2020 Nov 30;7:546779. doi: 10.3389/fsurg.2020.546779. eCollection 2020. Front Surg. 2020. PMID: 33425980 Free PMC article.
-
The influence of CI electrode array design on the preservation of residual hearing.Front Neurol. 2025 Jun 18;16:1599369. doi: 10.3389/fneur.2025.1599369. eCollection 2025. Front Neurol. 2025. PMID: 40606125 Free PMC article.
-
Postinsertional Cable Movements of Cochlear Implant Electrodes and Their Effects on Intracochlear Pressure.Biomed Res Int. 2016;2016:3937196. doi: 10.1155/2016/3937196. Epub 2016 Nov 9. Biomed Res Int. 2016. PMID: 27900329 Free PMC article.