A simple matrix of analytical performance to identify assays that risk patients using External Quality Assurance Program data
- PMID: 26800780
- DOI: 10.1016/j.clinbiochem.2016.01.014
A simple matrix of analytical performance to identify assays that risk patients using External Quality Assurance Program data
Abstract
Objectives: We propose a simple way to reliably rank assays for improvement according to patient risk, based solely on EQA imprecision and biological variation data. Because the underlying technique aligns the imprecision class of an assay from EQA data, peer performance can be used to assess achievable imprecision and the risk ranking can not only prioritise improvement but also highlight laboratory QC operating parameters that are easy to manage and provide reliable, acceptable performance.
Design and methods: A modified Failure Modes Effects Analysis (FMEA) is applied to produce an analyte risk rating based on three factors, each of which is graded: 1) the ease of detecting analytical errors based on the ratio of allowable limits of performance to imprecision (Assay Capability) compared to absolute standards and to peers, 2) the predicted frequency of errors in patient monitoring based on the ratio of within-individual biological variation to laboratory imprecision, and 3) the clinical importance of the assay as a surrogate marker for harm arising from an error.
Results: We provide laboratory examples to illustrate these models.
Conclusion: The proposed models using only EQA data can objectively identify assays at risk of failing against biological variation goals for monitoring patients and suggest parameters for reliable performance.
Keywords: Assay Capability; External quality assurance; FMEA; Quality Control; Risk.
Copyright © 2016 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Steady state errors and risk of a QC strategy.Clin Biochem. 2019 Feb;64:37-43. doi: 10.1016/j.clinbiochem.2018.12.005. Epub 2018 Dec 12. Clin Biochem. 2019. PMID: 30552866
-
Integrating quality control and external quality assurance.Clin Biochem. 2021 Sep;95:15-27. doi: 10.1016/j.clinbiochem.2021.05.003. Epub 2021 May 7. Clin Biochem. 2021. PMID: 33965412 Review.
-
Analytical performance specifications for external quality assessment - definitions and descriptions.Clin Chem Lab Med. 2017 Jun 27;55(7):949-955. doi: 10.1515/cclm-2017-0151. Clin Chem Lab Med. 2017. PMID: 28593915
-
[Domestic and international trends concerning allowable limits of error in external quality assessment scheme].Rinsho Byori. 2005 Jun;53(6):547-53. Rinsho Byori. 2005. PMID: 16026083 Japanese.
-
Reprint of "Influence of analytical bias and imprecision on the number of false positive results using Guideline-Driven Medical Decision Limits".Clin Chim Acta. 2014 May 15;432:127-34. doi: 10.1016/j.cca.2014.04.002. Epub 2014 Apr 18. Clin Chim Acta. 2014. PMID: 24747109 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical