. 2016 Mar;9(1):75-83.
doi: 10.1007/s12178-016-9322-2.

Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls

Affiliations

Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls

Mitchell C Weiser et al. Curr Rev Musculoskelet Med. 2016 Mar.

Abstract

The management of the patient with a recalled, modular neck-body total hip arthroplasty can be complex, as it involves a combination of clinical, technical, and medicolegal challenges. Management begins with a thorough history and physical exam, radiographic evaluation, infection workup, and serum metal ion levels. Three-dimensional imaging is obtained based on patient symptomatology and metal ion levels and is used to evaluate for the presence of an adverse local tissue response as well as the integrity of the existing soft tissue envelope. The decision to perform revision surgery is based on a combination of patient symptomatology, laboratory values, and imaging findings. Revision surgery involves the entire armamentarium of femoral revision techniques, and the acetabulum may need to be revised at the surgeon's discretion. The femoral implant can often be removed without disrupting the femoral bone envelope; however, the surgeon should have a low threshold to perform an extended trochanteric osteotomy.

Keywords: Adverse local tissue reaction; Corrosion; Modular neck; Revision hip arthroplasty.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
Preoperative and postoperative radiographs demonstrating a dual taper modular neck femoral stem revision to a titanium modular fluted tapered diaphyseal engaging stem utilizing the modified Steinman pin technique for stem removal

Similar articles

Cited by

References

    1. Older J. Charnley low-friction arthroplasty: a worldwide retrospective review at 15 to 20 years. J Arthroplasty. 2002;17(6):675–680. doi: 10.1054/arth.2002.31973. - DOI - PubMed
    1. Cooper HJ, Urban RM, Wixson RL, Meneghini RM, Jacobs JJ. Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual taper stem with a cobalt-chromium modular neck. J Bone Joint Surg Am. 2013;95(10):865–2. doi: 10.2106/JBJS.L.01042. - DOI - PMC - PubMed
    1. Molloy DO, Munir S, Jack CM, Cross MB, Walter WL, Walter WK., Sr Fretting and corrosion in modular-neck total hip arthroplasty femoral stems. J Bone Joint Surg Am. 2014;96(6):488–493. doi: 10.2106/JBJS.L.01625. - DOI - PubMed
    1. Meftah M, Haleem AM, Burn MB, Smith KM, Incavo SJ. Early corrosion-related failure of the rejuvenate modular total hip replacement system. J Bone Joint Surg Am. 2014;96(6):481–487. doi: 10.2106/JBJS.M.00979. - DOI - PubMed