Comparison of shear wave velocities on ultrasound elastography between different machines, transducers, and acquisition depths: a phantom study

Affiliations

Comparison of shear wave velocities on ultrasound elastography between different machines, transducers, and acquisition depths: a phantom study

Hyun Joo Shin et al. Eur Radiol. 2016 Oct.

Abstract

Objectives: To investigate consistency in shear wave velocities (SWVs) on ultrasound elastography using different machines, transducers and acquisition depths.

Methods: The SWVs were measured using an elasticity phantom with a Young's modulus of 16.9 kPa, with three recently introduced ultrasound elastography machines (A, B and C from different vendors) and two transducers (low and high frequencies) at four depths (2, 3, 4 and 5 cm). Mean SWVs from 15 measurements and coefficient of variations (CVs) were compared between three machines, two transducers and four acquisition depths.

Results: The SWVs using the high frequency transducer were not acquired at 5 cm depth in machine B, and a high frequency transducer was not available in machine C. The mean SWVs in the three machines were different (p ≤ 0.002). The CVs were 0-0.09 in three machines. The mean SWVs between the two transducers were different (p < 0.001) except at 4 and 5 cm depths in machine A. The SWVs were affected by the acquisition depths in all conditions (p < 0.001).

Conclusion: There is considerable difference in SWVs on ultrasound elastography depending on different machines, transducers and acquisition depths. Caution is needed when using the cutoff values of SWVs in different conditions.

Key points: • The shear wave velocities (SWVs) are different between different ultrasound elastography machines • The SWVs are also different between different transducers and acquisition depths • Caution is needed when using the cutoff SWVs measured under different conditions.

Keywords: Comparison; Consistency; Elastography; Phantom; Ultrasound.

PubMed Disclaimer

Similar articles

Cited by

References

    1. World J Gastroenterol. 2014 Jul 28;20(28):9578-84 - PubMed
    1. Ultrasonography. 2014 Jan;33(1):26-33 - PubMed
    1. J Ultrasound Med. 2012 Aug;31(8):1239-44 - PubMed
    1. Pediatr Radiol. 2015 Mar;45(3):376-85 - PubMed
    1. Ultrasonography. 2014 Jul;33(3):149-60 - PubMed

LinkOut - more resources