Flow-diverter devices in the treatment of intracranial aneurysms: A meta-analysis and systematic review
- PMID: 26838174
- PMCID: PMC4978339
- DOI: 10.1177/1971400915621321
Flow-diverter devices in the treatment of intracranial aneurysms: A meta-analysis and systematic review
Abstract
Objective: The purpose of this report was to discuss the overall limitations, safety and efficacy of flow-diverter stenting for intracranial aneurysms.
Methods: The authors performed a meta-analysis from January 2009 to September 2014 using the terms "flow diverter" and "intracranial aneurysms." Additional studies were identified through references in each reviewed article. Data extraction, performed independently by the authors, included demographic data, technical and clinical complications, morbidity and mortality, aneurismal occlusion rates related to flow-diverter devices. The analysis was performed using a fixed effect.
Results: Twenty-nine studies with 1524 patients and three to 62 months of follow-up were identified for analysis. The overall technical failure and complication rate was 9.3% (95% CI 6%-12.6%). The rate of procedure-related complication was 14% (95% CI 10.2%-17.9%) and 6.6% (95% CI 4%-9.1%) for morbidity and mortality. Fusiform, dissecting and circumferential aneurysm (OR 3.10, 95% CI 0.93-10.37) were significant risk factors for technical failure and complication. Posterior circulation location (OR 4.03, 95% CI 2.45-6.61), peripheral location (OR 2.74, 95% CI 1.52-4.94) and fusiform, dissecting and circumferential aneurysm (OR 1.95, 95% CI 1.15-3.30) were statistically significant risk factors for procedure-related complications. Posterior circulation location (OR 4.39, 95% CI 2.44-7.90) and peripheral location (OR 3.64, 95% CI 1.74-7.62) were statistically significant risk factors for morbidity and mortality.
Conclusions: Fusiform, dissecting and circumferential aneurysm, posterior circulation and peripheral locations have greater procedure-related complications.
Keywords: Intracranial aneurysm; endovascular treatment; flow-diverter stenting.
© The Author(s) 2016.
Figures
References
-
- Brinjikji W1, Murad MH, Lanzino G, et al. Endovascular treatment of intracranial aneurysms with flow diverters: A meta-analysis. Stroke 2013; 44: 442–447. - PubMed
-
- Arrese I, Sarabia R, Pintado R, et al. Flow-diverter devices for intracranial aneurysms: Systematic review and meta-analysis. Neurosurgery 2013; 73: 193–199. - PubMed
-
- Kocer N, Islak C, Kizilkilic O, et al. Flow re-direction endoluminal device in treatment of cerebral aneurysms: Initial experience with short-term follow-up results. J Neurosurg 2014; 120: 1158–1171. - PubMed
-
- Tan LA, Keigher KM, Munich SA, et al. Thromboembolic complications with Pipeline Embolization Device placement: Impact of procedure time, number of stents and pre-procedure P2Y12 reaction unit (PRU) value. J Neurointerv Surg 2015; 7: 217–221. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical