Abstract
Analysis of L-tryptophan (Trp) in biological samples has great importance for biomedical studies. Amino acid tyrosine (Tyr) that usually coexist with Trp in biological fluids can significantly interfere with reliable determination of Trp. In the current study, we demonstrate the development of two ultra-sensitive electrochemical sensor and label-free aptasensor for selective analysis of Trp in biological samples (i.e., cow's milk and human plasma, saliva and urine samples). In addition, without using AgCl/KCl, an Ag pseudo-reference screen printed electrode (Ag-PR-SPE) was exploited as a reference electrode. To prepare the engineered Trp sensor/aptasensor, a gold SPE was first modified with multiwall carbon nanotube (MWCNT-AuSPE) and then armed with Trp aptamer molecules (Apt-MWCNT-AuSPE). The prepared sensors were characterized using constant current-potentiometric stripping analysis (CC-PSA) and electrochemical impedance spectroscopy (EIS). The MWCNT-AuSPE and Apt-MWCNT-AuSPE were compared with respect to the linear detection range, limit of detection (LOD), accuracy, precision, repeatability. MWCNT-AuSPE and Apt-MWCNT-AuSPE demonstrate fast near-Nernstian response for PSA of Trp over the concentration ranging from 1.0 × 10(-9) to 2.0 × 10(-4) mol L(-1) and 1.0 × 10(-11) to 1.0 × 10(-4) mol L(-1) with detection limits of 3.6 × 10(-10) mol L(-1) and 4.9 × 10(-12) mol L(-1), respectively. Common interfering species present in the biological fluids (i.e., tyrosine, uric acid, ascorbic acid) showed no effects on the determination of Trp using CC-PSA. MWCNT-AuSPE and Apt-MWCNT-AuSPE represented well reproducibility and great precision with relative standard deviation (RSD) of 2.9% and 5.3% respectively. In comparison with the MWCNT-AuSPE, Apt-MWCNT-AuSPE provided higher sensitivity, selectivity and accuracy of Trp detection in real samples. Based on these findings, we propose the developed Apt-MWCNT-AuSPE as a simple detection method for analysis of Trp in biological samples.
Keywords: Multiwall carbon nanotube; Potentiometric stripping analysis; RNA aptamer; Screen-printed electrode.
Copyright © 2015 Elsevier B.V. All rights reserved.
Similar articles
-
Development of screen-printed tryptophan-kynurenine immunosensor for in vitro assay of kynurenine-mediated immunosuppression effect of cancer cells on activated T-cells.Biosens Bioelectron. 2017 Jun 15;92:287-293. doi: 10.1016/j.bios.2016.11.010. Epub 2016 Nov 6. Biosens Bioelectron. 2017. PMID: 27840038
-
Ultrasensitive and reusable electrochemical aptasensor for detection of tryptophan using of [Fe(bpy)3](p-CH3C6H4SO2)2 as an electroactive indicator.J Pharm Biomed Anal. 2019 Jan 30;163:180-187. doi: 10.1016/j.jpba.2018.10.006. Epub 2018 Oct 3. J Pharm Biomed Anal. 2019. PMID: 30316063
-
Construction of a highly sensitive signal-on aptasensor based on gold nanoparticles/functionalized silica nanoparticles for selective detection of tryptophan.Anal Bioanal Chem. 2017 Nov;409(27):6429-6438. doi: 10.1007/s00216-017-0588-z. Epub 2017 Aug 29. Anal Bioanal Chem. 2017. PMID: 28852807
-
A one-step aptasensor for ultrasensitive detection of lung cancer marker homocysteine based on multifunctional carbon nanotubes by square-wave voltammetry.Bioelectrochemistry. 2023 Oct;153:108464. doi: 10.1016/j.bioelechem.2023.108464. Epub 2023 May 15. Bioelectrochemistry. 2023. PMID: 37295310 Review.
-
Electrochemical aptasensors for contaminants detection in food and environment: Recent advances.Bioelectrochemistry. 2017 Dec;118:47-61. doi: 10.1016/j.bioelechem.2017.07.004. Epub 2017 Jul 12. Bioelectrochemistry. 2017. PMID: 28715665 Review.
Cited by
-
Malignancies and Biosensors: A Focus on Oral Cancer Detection through Salivary Biomarkers.Biosensors (Basel). 2021 Oct 15;11(10):396. doi: 10.3390/bios11100396. Biosensors (Basel). 2021. PMID: 34677352 Free PMC article. Review.
-
Highly Sensitive and Selective Detection of L-Tryptophan by ECL Using Boron-Doped Diamond Electrodes.Sensors (Basel). 2024 Jun 4;24(11):3627. doi: 10.3390/s24113627. Sensors (Basel). 2024. PMID: 38894416 Free PMC article.
-
Diagnosis of acute myocardial infarction: highlighting cardiac troponins as vital biomarkers.Bioimpacts. 2023;13(2):85-88. doi: 10.34172/bi.2023.22023. Epub 2022 Jan 16. Bioimpacts. 2023. PMID: 37193078 Free PMC article.
-
Advances in the design of nanomaterial-based electrochemical affinity and enzymatic biosensors for metabolic biomarkers: A review.Mikrochim Acta. 2018 May 2;185(5):276. doi: 10.1007/s00604-018-2820-8. Mikrochim Acta. 2018. PMID: 29721621 Review.
-