A Microscopic and Biomarker Evaluation of Embolic Filter Debris Collected During Carotid Artery Stenting
- PMID: 26839124
- DOI: 10.1177/1526602816628284
A Microscopic and Biomarker Evaluation of Embolic Filter Debris Collected During Carotid Artery Stenting
Abstract
Purpose: To evaluate and characterize debris retrieved from the cerebral embolic protection devices (EPDs) used during carotid artery stenting (CAS) and compare debris size, volume, tissue types, cellular composition, and protein biomarker expression in symptomatic and asymptomatic patients.
Methods: Distal protection filters were retrieved from 22 consecutive patients (mean age 71.6 years, range 52-85; 16 men) undergoing elective CAS between July 2012 and February 2014 for >70% internal carotid artery stenosis (mean 85.4% ± 10.3%). Six patients were symptomatic. The debris within each EPD was visually characterized using stereomicroscopy and then processed for histology and immunohistochemistry. Biomarkers were immunohistochemically measured to evaluate plaque stability [matrix metalloproteinase-9 (MMP-9)], inflammation [glycoprotein CD68 and interleukin-6 (IL-6)], or phenotype [smooth muscle (SM)-actin and type IV collagen]. The immunohistochemical results were measured using semiquantitative grading criteria based on both staining intensity and distribution in the samples.
Results: Macroscopic debris was visible in 5/22 EPDs; 3 of the 5 filters came from symptomatic patients. Microscopic debris was detected in all filters and ranged in size from 0.01 to 8.57 mm(2). Debris consisted of calcified, fibrous, and necrotic tissue, as well as fibrin and foam cells with no significant difference between the symptomatic and asymptomatic groups. There was no association between the degree or type of embolic material and stenosis severity, carotid tortuosity, calcium grade, soft plaque, or arch type. Symptomatic patients had a larger volume of debris (8.24 vs 0.58 mm(3), p<0.01), mean particle size (1.30 vs 0.32 mm(2), p<0.001), and expression of biomarkers IL-6 (2.17 vs 0.81, p<0.05), CD68 (2.00 vs 0.38, p<0.01), SM-actin (1.00 vs 0.25, p=0.055), type IV collagen (1.17 vs 0.25,p=0.082), and MMP-9 (1.00 vs 0.06, p<0.05).
Conclusion: Histological analysis revealed particulate embolization in all EPDs used during CAS. Symptomatic patients had a larger volume of embolic debris, mean particle size, and the biomarkers associated with inflammation, necrotic core, and diminished fibrous cap.
Keywords: angioplasty; atherosclerosis; biomarkers; carotid artery; carotid artery stenting; embolic protection device; filter; histology; immunohistochemistry; interleukin-6; matrix metalloproteinase-9; plaque; smooth muscle actin; stent.
© The Author(s) 2016.
Similar articles
-
Immunohistochemical Analysis of Debris Captured by Filter-Type Distal Embolic Protection Devices for Carotid Artery Stenting.J Stroke Cerebrovasc Dis. 2017 Apr;26(4):816-822. doi: 10.1016/j.jstrokecerebrovasdis.2016.10.025. Epub 2016 Nov 17. J Stroke Cerebrovasc Dis. 2017. PMID: 27865698
-
Quantitative analysis and predictors of embolic filter debris load during carotid artery stenting in asymptomatic patients.J Vasc Surg. 2018 Jul;68(1):109-117. doi: 10.1016/j.jvs.2017.09.055. Epub 2018 Mar 1. J Vasc Surg. 2018. PMID: 29503003
-
Cholesterol Crystals in Embolic Debris are Associated with Postoperative Cerebral Embolism after Carotid Artery Stenting.Cerebrovasc Dis. 2018;46(5-6):242-248. doi: 10.1159/000495795. Epub 2019 Jan 2. Cerebrovasc Dis. 2018. PMID: 30602147
-
Stenting for Internal Carotid Artery Stenosis Associated with Persistent Primitive Hypoglossal Artery Using Proximal Flow Blockade and Distal Protection System: A Technical Case Report and Literature Review.J Stroke Cerebrovasc Dis. 2016 Jun;25(6):e98-e102. doi: 10.1016/j.jstrokecerebrovasdis.2016.03.026. Epub 2016 Apr 19. J Stroke Cerebrovasc Dis. 2016. PMID: 27105567 Review.
-
Endarterectomy or carotid artery stenting: the quest continues.Am J Surg. 2008 Feb;195(2):259-69. doi: 10.1016/j.amjsurg.2007.07.022. Am J Surg. 2008. PMID: 18154764 Review.
Cited by
-
The relationship between ultrasonographic plaque morphology and histopathology of embolic material captured in the protective filter during carotid artery stenting.Interv Neuroradiol. 2023 Jun;29(3):285-290. doi: 10.1177/15910199221083117. Epub 2022 Mar 14. Interv Neuroradiol. 2023. PMID: 35285736 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous