Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Observational Study
. 2017 Jan;19(1):96-102.
doi: 10.1093/europace/euv410. Epub 2016 Feb 3.

Incidence and predictors of clinically relevant cardiac perforation associated with systematic implantation of active-fixation pacing and defibrillation leads: a single-centre experience with over 3800 implanted leads

Affiliations
Observational Study

Incidence and predictors of clinically relevant cardiac perforation associated with systematic implantation of active-fixation pacing and defibrillation leads: a single-centre experience with over 3800 implanted leads

Óscar Cano et al. Europace. 2017 Jan.

Abstract

Aims: Active-fixation leads have been associated with higher incidence of cardiac perforation. Large series specifically evaluating this complication are lacking. We sought to evaluate the incidence and predictors of clinically relevant cardiac perforation in a consecutive series of patients implanted with active-fixation pacing and defibrillation leads.

Methods and results: We conducted a retrospective observational study including all consecutive patients implanted with an active-fixation pacing/defibrillation lead at our institution from July 2008 to July 2015. The incidence of clinically relevant cardiac perforation and cardiac tamponade was evaluated. Univariate and multivariate analyses were used to identify predictors of cardiac perforation. Acute and long-term management of these patients was also investigated. A total of 3822 active-fixation pacing (n = 3035) and defibrillation (n = 787) leads were implanted in 2200 patients. Seventeen patients (0.8%) had clinically relevant cardiac perforation (13 acute and 4 subacute perforations), and 13 (0.5%) had cardiac tamponade resolved with pericardiocentesis. None of the patients with cardiac perforation required surgical treatment. In multivariate analysis, an age >80 years (OR 3.84, 95% CI 1.14-12.87, P = 0.029), female sex (OR 3.14, 95% CI 1.07-9.22, P = 0.037), and an apical position of the right ventricular lead (OR 3.37, 95% CI 1.17-9.67, P = 0.024) were independent predictors of cardiac perforation.

Conclusions: Implantation of active-fixation leads is associated with a low incidence of clinically relevant cardiac perforation. Older and female patients have a higher risk of perforation as well as those patients receiving the ventricular lead in an apical position.

Keywords: Active-fixation leads; Cardiac perforation; Cardiac tamponade; Implantable cardioverter defibrillator; Pacemaker.

PubMed Disclaimer

Publication types

MeSH terms