Ceramic Heads Decrease Metal Release Caused by Head-taper Fretting and Corrosion
- PMID: 26847452
- PMCID: PMC4773353
- DOI: 10.1007/s11999-015-4683-1
Ceramic Heads Decrease Metal Release Caused by Head-taper Fretting and Corrosion
Erratum in
-
Erratum to: Ceramic Heads Decrease Metal Release Caused by Head-taper Fretting and Corrosion.Clin Orthop Relat Res. 2016 May;474(5):1344. doi: 10.1007/s11999-016-4781-8. Clin Orthop Relat Res. 2016. PMID: 26956249 Free PMC article. No abstract available.
Abstract
Background: Metal release resulting from taper fretting and corrosion is a clinical concern, because wear and corrosion products may stimulate adverse local tissue reactions. Unimodular hip arthroplasties have a conical taper between the femoral head (head bore taper) and the femoral stem (stem cone taper). The use of ceramic heads has been suggested as a way of reducing the generation of wear and corrosion products from the head bore/stem cone taper junction. A previous semiquantitative study found that ceramic heads had less visual evidence of fretting-corrosion damage compared with CoCr heads; but, to our knowledge, no studies have quantified the volumetric material loss from the head bore and stem cone tapers of a matched cohort of ceramic and metal heads.
Questions/purposes: We asked: (1) Do ceramic heads result in less volume of material loss at the head-stem junction compared with CoCr heads; (2) do stem cone tapers have less volumetric material loss compared with CoCr head bore tapers; (3) do visual fretting-corrosion scores correlate with volumetric material loss; and (4) are device, patient, or intraoperative factors associated with volumetric material loss?
Methods: A quantitative method was developed to estimate volumetric material loss from the head and stem taper in previously matched cohorts of 50 ceramic and 50 CoCr head-stem pairs retrieved during revision surgery for causes not related to adverse reactions to metal particles. The cohorts were matched according to (1) implantation time, (2) stem flexural rigidity, and (3) lateral offset. Fretting corrosion was assessed visually using a previously published four-point, semiquantitative scoring system. The volumetric loss was measured using a precision roundness machine. Using 24 equally spaced axial traces, the volumetric loss was estimated using a linear least squares fit to interpolate the as-manufactured surfaces. The results of this analysis were considered in the context of device (taper angle clearance, head size, head offset, lateral offset, stem material, and stem surface finish) and patient factors that were obtained from the patients' operative records (implantation time, age at insertion, activity level, and BMI).
Results: The cumulative volumetric material losses estimated for the ceramic cohort had a median of 0.0 mm(3) per year (range, 0.0-0.4 mm(3)). The cumulative volumetric material losses estimated for the CoCr cohort had a median of 0.1 mm(3) per year (range, 0.0-8.8 mm(3)). An order of magnitude reduction in volumetric material loss was found when a ceramic head was used instead of a CoCr head (p < 0.0001). In the CoCr cohort, the femoral head bore tapers had a median material loss of 0.02 mm(3) (range, 0.0-8.7 mm(3)) and the stem cone tapers had a median material loss of 0.0 mm(3) (range, 0.0-0.32 mm(3)/year). There was greater material loss from femoral head bore tapers compared with stem cone tapers in the CoCr cohort (p < 0.001). There was a positive correlation between visual scoring and volumetric material loss (Spearman's ρ = 0.67, p < 0.01). Although visual scoring was effective for preliminary screening to separate tapers with no or mild damage from tapers with moderate to severe damage, it was not capable of discriminating in the large range of material loss observed at the taper surfaces with moderate to severe fretting-corrosion damage, indicated with a score of 3 or 4. We observed no correlations between volumetric material loss and device and patient factors.
Conclusions: The majority of estimated material loss from the head bore-stem cone junctions resulting from taper fretting and corrosion was from the CoCr head bore tapers as opposed to the stem cone tapers. Additionally, the total material loss from the ceramic cohort showed a reduction in the amount of metal released by an order of magnitude compared with the CoCr cohort.
Clinical relevance: We found that ceramic femoral heads may be an effective means by which to reduce metal release caused by taper fretting and corrosion at the head bore-stem cone modular interface in THAs.
Figures




Similar articles
-
Do ceramic femoral heads reduce taper fretting corrosion in hip arthroplasty? A retrieval study.Clin Orthop Relat Res. 2013 Oct;471(10):3270-82. doi: 10.1007/s11999-013-3096-2. Epub 2013 Jun 13. Clin Orthop Relat Res. 2013. PMID: 23761174 Free PMC article.
-
Otto Aufranc Award: Large Heads Do Not Increase Damage at the Head-neck Taper of Metal-on-polyethylene Total Hip Arthroplasties.Clin Orthop Relat Res. 2016 Feb;474(2):330-8. doi: 10.1007/s11999-015-4468-6. Clin Orthop Relat Res. 2016. PMID: 26208607 Free PMC article.
-
Does Surface Topography Play a Role in Taper Damage in Head-neck Modular Junctions?Clin Orthop Relat Res. 2016 Oct;474(10):2232-42. doi: 10.1007/s11999-016-4933-x. Clin Orthop Relat Res. 2016. PMID: 27339123 Free PMC article.
-
Evidence based recommendations for reducing head-neck taper connection fretting corrosion in hip replacement prostheses.Hip Int. 2017 Nov 21;27(6):523-531. doi: 10.5301/hipint.5000545. Epub 2017 Oct 9. Hip Int. 2017. PMID: 29027189 Review.
-
In Vivo Corrosion of Sleeved Ceramic Femoral Heads: A Retrieval Study.J Arthroplasty. 2021 Mar;36(3):1133-1137. doi: 10.1016/j.arth.2020.10.005. Epub 2020 Oct 9. J Arthroplasty. 2021. PMID: 33121849 Review.
Cited by
-
[Total hip arthroplasty in young patients : Bearings and custom-made prostheses].Orthopade. 2019 Apr;48(4):292-299. doi: 10.1007/s00132-019-03692-y. Orthopade. 2019. PMID: 30737518 Review. German.
-
Fretting and Fretting Corrosion Behavior of Additively Manufactured Ti-6Al-4V and Ti-Nb-Zr Alloys in Air and Physiological Solutions.J Funct Biomater. 2024 Feb 5;15(2):38. doi: 10.3390/jfb15020038. J Funct Biomater. 2024. PMID: 38391891 Free PMC article.
-
Pseudotumor in ceramic-on-metal total hip arthroplasty.Arthroplast Today. 2017 Oct 6;3(4):220-224. doi: 10.1016/j.artd.2017.09.001. eCollection 2017 Dec. Arthroplast Today. 2017. PMID: 29204484 Free PMC article.
-
Does increased diameter of metal femoral head associated with highly cross-linked polyethylene augment stress on the femoral stem and cortical hypertrophy?Int Orthop. 2021 May;45(5):1169-1177. doi: 10.1007/s00264-021-04994-7. Epub 2021 Feb 22. Int Orthop. 2021. PMID: 33619587
-
Variability in stem taper surface topography affects the degree of corrosion and fretting in total hip arthroplasty.Sci Rep. 2021 Apr 30;11(1):9348. doi: 10.1038/s41598-021-88234-3. Sci Rep. 2021. PMID: 33931680 Free PMC article.
References
-
- Arnholt C, Underwood R, MacDonald D, Higgs G, Chen A, Klein G, Hamlin B, Lee G, Mont M, Cates H, Malkani A, Kraay M, Rimnac C, Kurtz S. Micro-grooved Surface Topography Does Not Influence Fretting Corrosion of Tapers in THA: Classification and Retrieval Analysis. In: Greenwald AS, Kurtz SM, Lemons JE, Mihalko WM, editors. Modularity and Tapers in Total Joint Replacement Devices, STP 1591. Conshohocken, PA: American Society of Testing Materials (ASTM); 2015. pp. 99–112.
-
- Australian Orthopaedic Association National Joint Replacement Registry. Annual Report 2014. Available at: https://aoanjrr.dmac.adelaide.edu.au/en/annual-reports-2014. Accessed December 11, 2015.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical