Intra-coil interactions in split gradient coils in a hybrid MRI-LINAC system
- PMID: 26852418
- DOI: 10.1016/j.jmr.2016.01.013
Intra-coil interactions in split gradient coils in a hybrid MRI-LINAC system
Abstract
An MRI-LINAC system combines a magnetic resonance imaging (MRI) system with a medical linear accelerator (LINAC) to provide image-guided radiotherapy for targeting tumors in real-time. In an MRI-LINAC system, a set of split gradient coils is employed to produce orthogonal gradient fields for spatial signal encoding. Owing to this unconventional gradient configuration, eddy currents induced by switching gradient coils on and off may be of particular concern. It is expected that strong intra-coil interactions in the set will be present due to the constrained return paths, leading to potential degradation of the gradient field linearity and image distortion. In this study, a series of gradient coils with different track widths have been designed and analyzed to investigate the electromagnetic interactions between coils in a split gradient set. A driving current, with frequencies from 100 Hz to 10 kHz, was applied to study the inductive coupling effects with respect to conductor geometry and operating frequency. It was found that the eddy currents induced in the un-energized coils (hereby-referred to as passive coils) positively correlated with track width and frequency. The magnetic field induced by the eddy currents in the passive coils with wide tracks was several times larger than that induced by eddy currents in the cold shield of cryostat. The power loss in the passive coils increased with the track width. Therefore, intra-coil interactions should be included in the coil design and analysis process.
Keywords: Eddy currents; Electromagnetic interactions; Gradient coil; MRI–LINAC.
Copyright © 2016 Elsevier Inc. All rights reserved.
Similar articles
-
Mitigation of Intra-coil Eddy Currents in Split Gradient Coils in a Hybrid MRI-LINAC System.IEEE Trans Biomed Eng. 2017 Mar;64(3):725-732. doi: 10.1109/TBME.2016.2573316. Epub 2016 May 26. IEEE Trans Biomed Eng. 2017. PMID: 27249823
-
Skin and proximity effects in the conductors of split gradient coils for a hybrid Linac-MRI scanner.J Magn Reson. 2014 May;242:86-94. doi: 10.1016/j.jmr.2014.02.002. Epub 2014 Feb 12. J Magn Reson. 2014. PMID: 24607826
-
Longitudinal gradient coil optimization in the presence of transient eddy currents.Magn Reson Med. 2007 Jun;57(6):1119-30. doi: 10.1002/mrm.21243. Magn Reson Med. 2007. PMID: 17534904
-
Unconventional gradient coil designs in magnetic resonance imaging.Crit Rev Biomed Eng. 2014;42(6):493-526. doi: 10.1615/critrevbiomedeng.2015013100. Crit Rev Biomed Eng. 2014. PMID: 25955713 Review.
-
The Improvement of Flaw Detection by the Configuration of Uniform Eddy Current Probes.Sensors (Basel). 2019 Jan 18;19(2):397. doi: 10.3390/s19020397. Sensors (Basel). 2019. PMID: 30669390 Free PMC article. Review.
Cited by
-
Technical Note: Effects of rotating gantry on magnetic field and eddy currents in 0.35 T MRI-guided radiotherapy (MR-IGRT) system.Med Phys. 2021 Nov;48(11):7228-7235. doi: 10.1002/mp.15226. Epub 2021 Sep 25. Med Phys. 2021. PMID: 34520081 Free PMC article.
-
System-dependent image distortion related to gantry positions of a 0.35 T MRgRT: Characterization and the corresponding correction.J Appl Clin Med Phys. 2023 Jan;24(1):e13826. doi: 10.1002/acm2.13826. Epub 2022 Nov 10. J Appl Clin Med Phys. 2023. PMID: 36354747 Free PMC article.
-
Design and construction of gradient coils for an MRI-guided small animal radiation platform.Heliyon. 2024 Feb 19;10(4):e26251. doi: 10.1016/j.heliyon.2024.e26251. eCollection 2024 Feb 29. Heliyon. 2024. PMID: 39670066 Free PMC article.
Publication types
LinkOut - more resources
Full Text Sources
Other Literature Sources