Effects of Pump Pulsation on Hydrodynamic Properties and Dissolution Profiles in Flow-Through Dissolution Systems (USP 4)
- PMID: 26869175
- DOI: 10.1007/s11095-016-1874-8
Effects of Pump Pulsation on Hydrodynamic Properties and Dissolution Profiles in Flow-Through Dissolution Systems (USP 4)
Abstract
Purpose: To clarify the effects of pump pulsation and flow-through cell (FTC) dissolution system settings on the hydrodynamic properties and dissolution profiles of model formulations.
Methods: Two FTC systems with different cell temperature control mechanisms were used. Particle image velocimetry (PIV) was used to analyze the hydrodynamic properties of test solutions in the flow-through dissolution test cell. Two pulsation pumps (semi-sine, full-sine) and a non-pulsatile pump were used to study the effects of varied flows on the dissolution profiles of United States Pharmacopeia standard tablets.
Results: PIV analysis showed periodic changes in the aligned upward fluid flow throughout the dissolution cell that was designed to reduce the temperature gradient during pump pulsation (0.5 s/pulse). The maximum instantaneous flow from the semi-sine pump was higher than that of the full-sine pump under all conditions. The flow from the semi-sine wave pump showed faster dissolution of salicylic acid and prednisone tablets than those from other pumps. The semi-sine wave pump flow showed similar dissolution profiles in the two FTC systems.
Conclusions: Variations in instantaneous fluid flow caused by pump pulsation that meets the requirements of pharmacopoeias are a factor that affects the dissolution profiles of tablets in FTC systems.
Keywords: dissolution testing; flow-through cell system; hydrodynamics; pulsatile pump.
Similar articles
-
Altered Media Flow and Tablet Position as Factors of How Air Bubbles Affect Dissolution of Disintegrating and Non-disintegrating Tablets Using a USP 4 Flow-Through Cell Apparatus.AAPS PharmSciTech. 2021 Aug 24;22(7):227. doi: 10.1208/s12249-021-02117-4. AAPS PharmSciTech. 2021. PMID: 34431011
-
Particle Image Velocimetry Evaluation of Fluid Flow Profiles in USP 4 Flow-Through Dissolution Cells.Pharm Res. 2015 Sep;32(9):2950-9. doi: 10.1007/s11095-015-1676-4. Epub 2015 Mar 20. Pharm Res. 2015. PMID: 25791217
-
Hydrodynamic, mass transfer, and dissolution effects induced by tablet location during dissolution testing.J Pharm Sci. 2009 Apr;98(4):1511-31. doi: 10.1002/jps.21512. J Pharm Sci. 2009. PMID: 18781589
-
Effect of basket mesh size on the hydrodynamics of a partially filled (500 mL) USP rotating basket dissolution testing Apparatus 1.Int J Pharm. 2024 Jun 10;658:124209. doi: 10.1016/j.ijpharm.2024.124209. Epub 2024 May 6. Int J Pharm. 2024. PMID: 38718973
-
Engineering tools for understanding the hydrodynamics of dissolution tests.Drug Dev Ind Pharm. 2003 Feb;29(2):231-9. doi: 10.1081/ddc-120016731. Drug Dev Ind Pharm. 2003. PMID: 12648020 Review.
Cited by
-
Insights into the Mechanism of Enhanced Dissolution in Solid Crystalline Formulations.Pharmaceutics. 2024 Apr 7;16(4):510. doi: 10.3390/pharmaceutics16040510. Pharmaceutics. 2024. PMID: 38675170 Free PMC article.
-
Effects of Glass Bead Size on Dissolution Profiles in Flow-through Dissolution Systems (USP 4).AAPS PharmSciTech. 2024 Oct 22;25(8):251. doi: 10.1208/s12249-024-02972-x. AAPS PharmSciTech. 2024. PMID: 39433646
-
Altered Media Flow and Tablet Position as Factors of How Air Bubbles Affect Dissolution of Disintegrating and Non-disintegrating Tablets Using a USP 4 Flow-Through Cell Apparatus.AAPS PharmSciTech. 2021 Aug 24;22(7):227. doi: 10.1208/s12249-021-02117-4. AAPS PharmSciTech. 2021. PMID: 34431011
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources