An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files
- PMID: 26877985
- PMCID: PMC4751208
- DOI: 10.5395/rde.2016.41.1.6
An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files
Abstract
Objectives: The purpose of this in vitro study was to evaluate the accuracy of working length (WL) determination of four electronic apex locators (EALs), namely, Root ZX (RZX), Elements diagnostic unit and apex locator (ELE), SybronEndo Mini Apex locator (MINI) and Propex pixi (PIXI) using Stainless steel (SS) and nickel-titanium (NiTi) hand files. The null hypothesis was that there was no difference between canal length determination by SS and NiTi files of 4 EALs.
Materials and methods: Sixty extracted, single rooted human teeth were decoronated and the canal orifice flared. The actual length (AL) was assessed visually, and the teeth were embedded in an alginate model. The electronic length (EL) measurements were recorded with all four EALs using SS and NiTi files at '0.5' reading on display. The differences between the AL and EL were compared.
Results: The results obtained with each EAL with SS and NiTi files were compared with AL. A paired sample t test showed that there was a statistical significant difference between EAL readings with SS and NiTi files for RZX and MINI (p < 0.05). The accuracy of RZX, ELE, MINI and PIXI within ± 0.5 mm of AL with SS/NiTi files were 93.3%/70%, 90%/91.7%, 95%/68.3%, and 83.3%/83.3%, respectively.
Conclusions: The results of this study indicate that Root ZX was statistically more accurate with NiTi files compared to SS files, while MINI was statistically more accurate with SS files compared to NiTi files. ELE and PIXI were not affected by the alloy type of the file used to determine WL.
Keywords: Electronic apex locator; Elements Apex locator; Mini Apex Locator; NiTi hand files; Propex pixi; Root ZX.
Conflict of interest statement
Similar articles
-
An in vivo comparison of accuracy of two electronic apex locators in determining working length using stainless steel and nickel titanium files.Clin Cosmet Investig Dent. 2018 May 22;10:75-82. doi: 10.2147/CCIDE.S158882. eCollection 2018. Clin Cosmet Investig Dent. 2018. PMID: 29861643 Free PMC article.
-
An in vivo comparison of the Root ZX II, the Apex NRG XFR, and Mini Apex Locator by using rotary nickel-titanium files.J Endod. 2009 Jul;35(7):962-5. doi: 10.1016/j.joen.2009.04.025. J Endod. 2009. PMID: 19567315
-
In Vivo Evaluation of 3 Electronic Apex Locators: Root ZX Mini, Apex ID, and Propex Pixi.J Endod. 2020 Feb;46(2):158-161. doi: 10.1016/j.joen.2019.10.035. Epub 2019 Dec 12. J Endod. 2020. PMID: 31839411
-
Influence of Coronal Preflaring on the Accuracy of Electronic Working Length Determination: Systematic Review and Meta-Analysis.J Clin Med. 2021 Jun 23;10(13):2760. doi: 10.3390/jcm10132760. J Clin Med. 2021. PMID: 34201705 Free PMC article. Review.
-
Accuracy of different generations of apex locators in determining working length; a systematic review and meta-analysis.Saudi Dent J. 2022 Jan;34(1):11-20. doi: 10.1016/j.sdentj.2021.09.020. Epub 2021 Sep 25. Saudi Dent J. 2022. PMID: 35068894 Free PMC article. Review.
Cited by
-
Factors influencing the accuracy of electronic apex locators: A scoping review.Dent Res J (Isfahan). 2025 Jun 27;22:26. doi: 10.4103/drj.drj_634_23. eCollection 2025. Dent Res J (Isfahan). 2025. PMID: 40655901 Free PMC article. Review.
-
An in vivo comparison of accuracy of two electronic apex locators in determining working length using stainless steel and nickel titanium files.Clin Cosmet Investig Dent. 2018 May 22;10:75-82. doi: 10.2147/CCIDE.S158882. eCollection 2018. Clin Cosmet Investig Dent. 2018. PMID: 29861643 Free PMC article.
-
The Accuracy of Endodontic Length Measurement Using Cone-Beam Computed Tomography in Comparison with Electronic Apex Locators.Iran Endod J. 2020 Winter;15(1):12-17. doi: 10.22037/iej.v15i1.26720. Iran Endod J. 2020. PMID: 36704319 Free PMC article.
-
The Precision of Propex Pixi with Different Instruments and Coronal Preflaring Procedures.Eur Endod J. 2019 Jul 10;4(2):75-79. doi: 10.14744/eej.2019.52724. eCollection 2019. Eur Endod J. 2019. PMID: 32161891 Free PMC article.
-
Influence of heat treatment of nickel-titanium instruments on the accuracy of an electronic apex locator integrated with endodontic motor.J Conserv Dent. 2022 Nov-Dec;25(6):596-599. doi: 10.4103/jcd.jcd_260_22. Epub 2022 Aug 17. J Conserv Dent. 2022. PMID: 36591588 Free PMC article.
References
-
- Ingle JI, Backland LK. Endodontics. 5th ed. Loma Linda, California: BC Decker Inc.; 2002. Chapter 10. Endodontic cavity preparation; pp. 510–525.
-
- Kim E, Lee SJ. Electronic apex locator. Dent Clin North Am. 2004;48:35–54. - PubMed
-
- Sunada I. New method for measuring the length of the root canal. J Dent Res. 1962;41:375–387.
-
- Nekoofar MH, Ghandi MM, Hayes SJ, Dummer PM. The fundamental operating principles of electronic root canal length measurement devices. Int Endod J. 2006;39:595–609. - PubMed
-
- Kobayashi C, Suda H. New electronic canal measuring device based on the ratio method. J Endod. 1994;20:111–114. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous