Multicenter Study
. 2016 Feb 22;9(4):367-372.
doi: 10.1016/j.jcin.2015.12.012.

Transcatheter Aortic Valve Replacement Using the Repositionable LOTUS Valve: United Kingdom Experience

Affiliations
Free article
Multicenter Study

Transcatheter Aortic Valve Replacement Using the Repositionable LOTUS Valve: United Kingdom Experience

Rajiv Rampat et al. JACC Cardiovasc Interv. .
Free article

Abstract

Objectives: This study sought to present the U.K. experience to date with the second-generation LOTUS bioprosthesis (Boston Scientific, Natick, Massachusetts).

Background: First-generation transcatheter aortic valves have limitations. Second-generation repositionable valves may improve on some of those limitations.

Methods: Prospectively collected data relating to procedural and in-hospital outcome was analyzed from 10 implantation centers in the United Kingdom.

Results: Implants in 228 patients age 81.4 ± 7.6 years were studied; 53.5% were male. Mean logistic EuroScore was 17.5 ± 12.4. One hundred eighty-seven (82.0%) were undertaken for aortic stenosis, 7 (3.1%) for aortic regurgitation, and 34 (14.9%) for mixed aortic valve disease. A total of 67.1% of cases were done under local anesthetic and/or sedation with transfemoral access in 94.7% and transaortic in 5.3%. Three device sizes were used: 23 mm (n = 66, 28.9%), 25 mm (n = 39, 17.1%), and 27 mm (n = 123, 53.9%). The valve was successfully deployed in 99.1% of procedures. After implantation, the mean aortic gradient was 11.4 ± 5.4 mm Hg and aortic valve area 1.6 ± 0.5 cm(2). In-hospital mortality was 1.8% (n = 4). Complications included cardiac tamponade (1.8%), conversion to sternotomy (1.3%), stroke (3.9%), vascular access-related (7.0%), and acute kidney injury (7.9%). The incidence of moderate/severe aortic regurgitation was 0.8% (n = 2). A total of 31.8% of patients required new permanent pacemaker implantation.

Conclusions: This analysis represents the largest published series on use of the LOTUS valve. Outcomes using this valve are excellent. In-hospital mortality is very low. Complication rates are low, and the LOTUS valve improves on first-generation valves, particularly with regard to residual aortic regurgitation.

Keywords: LOTUS; aortic stenosis; complication; outcome; transcatheter aortic valve.