Comparative Study
. 2017 Jan;19(1):81-87.
doi: 10.1093/europace/euv455. Epub 2016 Feb 17.

Implantable cardioverter defibrillator therapy in young individuals: comparison of conventional and subcostal approaches-a single-centre experience

Affiliations
Comparative Study

Implantable cardioverter defibrillator therapy in young individuals: comparison of conventional and subcostal approaches-a single-centre experience

Ghazala Irfan et al. Europace. 2017 Jan.

Abstract

Aim: The aim of our study is to compare two approaches of implantable cardiac defibrillator (ICD) implantation, conventional (supra/subpectoral) and subcostal in young adults in terms of procedural complications and adverse events encountered during follow-up.

Methods and results: From January 2007 to December 2013, all patients under the age of 50 years who received an ICD in our centre were included in this study. Patient's hospital records were analysed for procedural complications and adverse events during follow-up until December 2014. Data from device on first interrogation after implantation and on follow-up were also noted. A total of 106 patients of which 40.6% had Brugada's syndrome (65.1% male, age 33.6 ± 10.97 years) were included in analysis; 71 (61%) had ICD placed in (sub/supra) pectoral and 35 (33%) in subcostal position. Only seven patients received an epicardial lead system. During the follow-up period of 2.1 ± 1.8 years, 84.90% of the patients had no adverse events. Most of the complications, procedural and during follow-up, occur in conventionally placed, pectoral ICD. Lead follow-up data in both groups, conventional and subcostal, showed no difference in right ventricular (RV) shock impedance and R wave sensing, P-value = 0.56 and 0.77, respectively. Lead survival was 95 and 97%, respectively, in conventional and subcostal groups over a mean follow-up of 2.1 ± 1.8 years. Log-rank test for lead survival was not significant in terms of site of implantation.

Conclusion: To the best our knowledge, this is the first study demonstrating subcostal ICD placement in young adults and resulting in equivalent to better outcomes when compared with conventionally placed pectoral ICD. Subcostal ICD placement might be considered an alternative option in young adults as it results in better procedural outcomes and also comparable rate of adverse events during follow-up, but bigger studies with a larger number of patients are needed for a definitive conclusion.

Keywords: Cardiac device complications; Implantable cardioverter defibrillator; Implantation technique; Lead failure; Subcostal ICD.