Comparative Study
. 2016 Jun;254(6):1175-80.
doi: 10.1007/s00417-016-3291-4. Epub 2016 Mar 1.

Visual acuity measured with a smartphone app is more accurate than Snellen testing by emergency department providers

Affiliations
Comparative Study

Visual acuity measured with a smartphone app is more accurate than Snellen testing by emergency department providers

Akhilesh S Pathipati et al. Graefes Arch Clin Exp Ophthalmol. 2016 Jun.

Abstract

Purpose: To assess the accuracy of best-corrected visual acuity (BCVA) measured by non-ophthalmic emergency department (ED) staff with a standard Snellen chart versus an automated application (app) on a handheld smartphone (Paxos Checkup, San Francisco, CA, USA).

Methods: The study included 128 subjects who presented to the Stanford Hospital ED for whom the ED requested an ophthalmology consultation. We conducted the study in two phases. During phase 1 of the study, ED staff tested patient BCVA using a standard Snellen test at 20 feet. During phase 2 of the study, ED staff tested patient near BCVA using the app. During both phases, ophthalmologists measured BCVA with a Rosenbaum near chart, which was treated as the gold standard. ED BCVA measurements were benchmarked prospectively against ophthalmologists' measurements and converted to logMAR.

Results: ED logMAR BCVA was 0.21 ± 0.35 (approximately 2 Snellen lines difference ± 3 Snellen lines) higher than that of ophthalmologists when ED staff used a Snellen chart (p = .0.00003). ED BCVA was 0.06 ± 0.40 (less than 1 Snellen line ± 4 Snellen lines) higher when ED staff used the app (p = 0.246). Inter-observer difference was therefore smaller by more than 1 line (0.15 logMAR) with the app (p = 0.046).

Conclusions: BCVA measured by non-ophthalmic ED staff with an app was more accurate than with a Snellen chart. Automated apps may provide a means to standardize and improve the efficiency of ED ophthalmologic care.

Keywords: Emergency department; Telemedicine; Vision testing; Visual acuity.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Arch Ophthalmol. 2002 Jun;120(6):774-80 - PubMed
    1. J Telemed Telecare. 2013 Jan;19(1):55-9 - PubMed
    1. BMJ Open. 2013 Jun 20;3(6):null - PubMed
    1. J Cataract Refract Surg. 2004 Feb;30(2):287-90 - PubMed
    1. J Emerg Nurs. 2008 Jun;34(3):233-5 - PubMed

Publication types

LinkOut - more resources