Assessment of reliability of CAD-CAM tooth-colored implant custom abutments
- PMID: 26948082
- DOI: 10.1016/j.prosdent.2015.12.015
Assessment of reliability of CAD-CAM tooth-colored implant custom abutments
Abstract
Statement of problem: Information is lacking about the fatigue resistance of computer-aided design and computer-aided manufacturing (CAD-CAM) tooth-colored implant custom abutment materials.
Purpose: The purpose of this in vitro study was to investigate the reliability of different types of CAD-CAM tooth-colored implant custom abutments.
Material and methods: Zirconia (Lava Plus), lithium disilicate (IPS e.max CAD), and resin-based composite (Lava Ultimate) abutments were fabricated using CAD-CAM technology and bonded to machined titanium-6 aluminum-4 vanadium (Ti-6Al-4V) alloy inserts for conical connection implants (NobelReplace Conical Connection RP 4.3×10 mm; Nobel Biocare). Three groups (n=19) were assessed: group ZR, CAD-CAM zirconia/Ti-6Al-4V bonded abutments; group RC, CAD-CAM resin-based composite/Ti-6Al-4V bonded abutments; and group LD, CAD-CAM lithium disilicate/Ti-6Al-4V bonded abutments. Fifty-seven implant abutments were secured to implants and embedded in autopolymerizing acrylic resin according to ISO standard 14801. Static failure load (n=5) and fatigue failure load (n=14) were tested. Weibull cumulative damage analysis was used to calculate step-stress reliability at 150-N and 200-N loads with 2-sided 90% confidence limits. Representative fractured specimens were examined using stereomicroscopy and scanning electron microscopy to observe fracture patterns.
Results: Weibull plots revealed β values of 2.59 for group ZR, 0.30 for group RC, and 0.58 for group LD, indicating a wear-out or cumulative fatigue pattern for group ZR and load as the failure accelerating factor for groups RC and LD. Fractographic observation disclosed that failures initiated in the interproximal area where the lingual tensile stresses meet the compressive facial stresses for the early failure specimens. Plastic deformation of titanium inserts with fracture was observed for zirconia abutments in fatigue resistance testing.
Conclusions: Significantly higher reliability was found in group ZR, and no significant differences in reliability were determined between groups RC and LD. Differences were found in the failure characteristics of group ZR between static and fatigue loading.
Copyright © 2016 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights reserved.
Similar articles
-
In vitro assessment of three types of zirconia implant abutments under static load.J Prosthet Dent. 2013 Apr;109(4):255-63. doi: 10.1016/S0022-3913(13)60054-2. J Prosthet Dent. 2013. PMID: 23566607
-
Comparison of fracture resistance of pressable metal ceramic custom implant abutment with a commercially fabricated CAD/CAM zirconia implant abutment.J Prosthet Dent. 2013 Nov;110(5):389-96. doi: 10.1016/j.prosdent.2013.06.004. Epub 2013 Sep 5. J Prosthet Dent. 2013. PMID: 24011801
-
Fracture resistance of titanium and zirconia abutments: an in vitro study.J Prosthet Dent. 2013 May;109(5):304-12. doi: 10.1016/S0022-3913(13)60306-6. J Prosthet Dent. 2013. PMID: 23684280 Clinical Trial.
-
Fits of Implant Zirconia Custom Abutments and Frameworks: A Systematic Review and Meta-Analyses.Int J Oral Maxillofac Implants. 2019 January/February;34(1):99–114. doi: 10.11607/jomi.6657. Epub 2018 Oct 3. Int J Oral Maxillofac Implants. 2019. PMID: 30282084
-
Relation of CAD/CAM zirconia dental implant abutments with periodontal health and final aesthetic aspects; A systematic review.J Clin Exp Dent. 2023 Jan 1;15(1):e64-e70. doi: 10.4317/jced.59878. eCollection 2023 Jan. J Clin Exp Dent. 2023. PMID: 36755676 Free PMC article. Review.
Cited by
-
Tensile Bond Strength of Three Custom-made Tooth-Colored Implant Superstructures to Titanium Inserts.J Dent (Shiraz). 2019 Jun;20(2):131-136. doi: 10.30476/DENTJODS.2019.44924. J Dent (Shiraz). 2019. PMID: 31214642 Free PMC article.
-
Mechanical behavior of hybrid custom implant abutments with various crown materials: a 3D finite element analysis.BMC Oral Health. 2025 Jul 5;25(1):1106. doi: 10.1186/s12903-025-06445-w. BMC Oral Health. 2025. PMID: 40618097 Free PMC article.
-
In Vitro Assessment of a New Block Design for Implant Crowns with Functional Gradient Fabricated with Resin Composite and Zirconia Insert.Materials (Basel). 2024 Aug 2;17(15):3815. doi: 10.3390/ma17153815. Materials (Basel). 2024. PMID: 39124479 Free PMC article.
-
Biocompatibility and antibacterial properties of zirconium nitride coating on titanium abutments: An in vitro study.PLoS One. 2018 Jun 26;13(6):e0199591. doi: 10.1371/journal.pone.0199591. eCollection 2018. PLoS One. 2018. PMID: 29944716 Free PMC article.
-
A Comparative Study on the Mechanical Properties of a Polymer-Infiltrated Ceramic-Network Material Used for the Fabrication of Hybrid Abutment.Materials (Basel). 2018 Sep 11;11(9):1681. doi: 10.3390/ma11091681. Materials (Basel). 2018. PMID: 30208618 Free PMC article.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials
Miscellaneous