The applicability of correction factor for corneal thickness on non-contact tonometer measured intraocular pressure in LASIK treated eyes
- PMID: 26949354
- PMCID: PMC4759510
- DOI: 10.1016/j.sjopt.2015.11.001
The applicability of correction factor for corneal thickness on non-contact tonometer measured intraocular pressure in LASIK treated eyes
Abstract
Purpose: To determine the applicability of central corneal thickness (CCT) based correction factor for non-contact tonometer (NCT) measured intraocular pressure (IOP) readings.
Method: A prospective, non-randomized study involved 346 eyes of 173 consecutive patients with age ⩾21 years undergoing laser-assisted in situ keratomileusis (LASIK) for myopia and/or myopic astigmatism. The CCT and IOP were measured before and after the LASIK procedure. The IOP pre and post-LASIK was compared after applying the correction factor for CCT. Patients not completing the 3 month postoperative follow-up were excluded.
Results: The median spherical equivalent before undergoing LASIK was -4.25D (inter-quartile range, -3.25D). The mean preoperative CCT was 536.82 ± 33.71 μm which reduced to 477.55 ± 39.3 μm (p < 0.01) post-LASIK. The mean IOP reduced from a preoperative value of 14.6 ± 2.32 mmHg to 10.64 ± 2.45 mmHg postoperatively (p < 0.01). On applying correction for the corneal thickness, the pre and postoperative IOP was 15.14 ± 2.8 mmHg and 15.37 ± 2.65 mmHg (p = 0.06) respectively with a strong positive correlation (r = 0.7, p < 0.01). Three hundred eyes (86.7%) had an absolute difference in IOP of less than 3.0 mmHg post-CCT correction which is within the retest variability of NCT. Only 46 eyes (13.3%) had an absolute difference of more than 3.0 mmHg.
Conclusion: The modified Ehler's correction algorithm used in this study can be effectively applied in the normal IOP range in a majority of patients.
Keywords: CCT; Correction; Intraocular pressure; LASIK; NCT.
Figures
References
-
- Yuen L.H., Chan W.K., Koh J., Mehta J.S., Tan D.T., SingLasik Research Group A 10-year prospective audit of LASIK outcomes for myopia in 37,932 eyes at a single institution in Asia. Ophthalmology. 2010;117(1236–44):e1. - PubMed
-
- Rivera J.L., Bell N.P., Feldman R.M. Risk factors for primary open angle glaucoma progression: what we know and what we need to know. Curr Opin Ophthalmol. 2008;19:102–106. - PubMed
-
- Faucher A., Grégoire J., Blondeau P. Accuracy of Goldmann tonometry after refractive surgery. J Cataract Refract Surg. 1997;23:832–838. - PubMed
-
- Bochmann F., Ang G.S., Azuara-Blanco A. Lower corneal hysteresis in glaucoma patients with acquired pit of the optic nerve (APON) Graefes Arch Clin Exp Ophthalmol. 2008;246:735–738. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources