Abstract
Stent struts protruding into ostial side branch called "jailed strut" at bifurcation lesions is a likely cause of thrombus formation. We aimed to investigate the influences of multiple kissing balloon inflation (KBI) for stent expansion, and stent platform design, respectively, on the reduction of incomplete stent apposition area (ISA area) caused by jailed struts at a side-branch ostium, using a three-dimensional elastic left main (LM) bifurcated coronary artery model. The referenced LM bifurcation angle data of 209 patients were stratified by tertiles focusing on the angle between the LM trunk (LMT) and left anterior descending artery (LAD). A bifurcation model was fabricated with angles of 129°, 122.2°, and 76.4° for LMT-LAD, LMT-left circumflex (LCx), and LAD-LCx, respectively, and with diameters of 5, 3.75, and 3.5 mm for LMT, LAD, and LCx, respectively; these diameters fulfill Murray's law. A 75 % stenosis was included along the LMT. One-time and three-time KBIs were conducted using two-link Nobori and three-link Xience Xpedition (n = 6 each). The ISA area was quantified using micro-CT. Three-time KBI was effective in reducing the ISA area compared with one-time KBI for both the Nobori (p = 0.05) and Xience Xpedition (p = 0.07). The ISA area was smaller in the Nobori than in the Xience Xpedition, both in one-time and three-time KBI (one-time KBI: p = 0.003; three-time KBI: p = 0.001). Our findings of this study on reducing the ISA area by focusing on an interventional technique and stent design may help to improve coronary bifurcation intervention for a possibly better long-term clinical outcome.
Keywords: Incomplete stent apposition; Jailed strut; Left main bifurcation; Stent design; Stent expansion method.
Conflict of interest statement
Compliance with ethical standards Human rights statement This study was approved by the ethics committee of Shin-Koga Hospital, and all patients provided written informed consent. Conflict of interest This study was supported by a Health Science Research Grant (H20–IKOU-IPAN-001) from the Ministry of Health, Labour and Welfare, Japan. All authors have no conflict of interest. Funding source Health Science Research Grant from Ministry of Health, Labour and Welfare, Japan.
Figures





Similar articles
-
Differences of side branch jailing between left main-left anterior descending artery stenting and left main-left circumflex artery stenting with Nobori biolimus-eluting stent.Heart Vessels. 2016 Dec;31(12):1895-1903. doi: 10.1007/s00380-016-0812-3. Epub 2016 Feb 15. Heart Vessels. 2016. PMID: 26879743
-
Impact of kissing balloon inflation on the main vessel stent volume, area, and symmetry after side-branch dilation in patients with coronary bifurcation lesions: a serial volumetric intravascular ultrasound study.JACC Cardiovasc Interv. 2013 Sep;6(9):923-31. doi: 10.1016/j.jcin.2013.04.019. Epub 2013 Aug 14. JACC Cardiovasc Interv. 2013. PMID: 23954062
-
Comparative Analysis of Sequential Proximal Optimizing Technique Versus Kissing Balloon Inflation Technique in Provisional Bifurcation Stenting: Fractal Coronary Bifurcation Bench Test.JACC Cardiovasc Interv. 2015 Aug 24;8(10):1308-1317. doi: 10.1016/j.jcin.2015.05.016. JACC Cardiovasc Interv. 2015. PMID: 26315733
-
Stenting of bifurcation lesions: a rational approach.J Interv Cardiol. 2001 Dec;14(6):573-85. doi: 10.1111/j.1540-8183.2001.tb00375.x. J Interv Cardiol. 2001. PMID: 12053378 Review.
-
Coronary bifurcation lesions: Present status and future perspectives.Int J Cardiol. 2015;187:48-57. doi: 10.1016/j.ijcard.2015.03.183. Epub 2015 Mar 18. Int J Cardiol. 2015. PMID: 25828312 Review.
Cited by
-
Finite element analysis of cutting balloon expansion in a calcified artery model of circular angle 180°: Effects of balloon-to-diameter ratio and number of blades facing calcification on potential calcification fracturing and perforation reduction.PLoS One. 2021 May 13;16(5):e0251404. doi: 10.1371/journal.pone.0251404. eCollection 2021. PLoS One. 2021. PMID: 33984003 Free PMC article.
-
A Hypothetical Vascular Stent with Locally Enlarged Segment and the Hemodynamic Evaluation.Cardiol Res Pract. 2020 Feb 25;2020:7041284. doi: 10.1155/2020/7041284. eCollection 2020. Cardiol Res Pract. 2020. PMID: 32185077 Free PMC article.
-
How to use three-dimensional optical coherence tomography effectively in coronary bifurcation stenting.Front Cardiovasc Med. 2022 Nov 11;9:1023834. doi: 10.3389/fcvm.2022.1023834. eCollection 2022. Front Cardiovasc Med. 2022. PMID: 36440010 Free PMC article. Review.
-
Importance of optimal rewiring guided by 3-dimensional optical frequency domain imaging during double-kissing culotte stenting demonstrated through a novel bench model.Sci Rep. 2023 Aug 19;13(1):13511. doi: 10.1038/s41598-023-40606-7. Sci Rep. 2023. PMID: 37598246 Free PMC article.
-
Comparison of diagnostic performance in assessing the rewiring position into a jailed side branch between online 3D reconstruction systems version 1.1 and 1.2 derived from optical frequency domain imaging.Cardiovasc Interv Ther. 2020 Oct;35(4):336-342. doi: 10.1007/s12928-019-00629-2. Epub 2019 Nov 9. Cardiovasc Interv Ther. 2020. PMID: 31707692 Free PMC article.
References
-
- Hildick-Smith D, de Belder AJ, Cooter N, Curzen NP, Clayton TC, Oldroyd KG, et al. Randomized trial of simple versus complex drug-eluting stenting for bifurcation lesions: the British Bifurcation Coronary Study: old, new, and evolving strategies. Circulation. 2010;121:1235–1243. doi: 10.1161/CIRCULATIONAHA.109.888297. - DOI - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources