Patient movement and motion artefacts in cone beam computed tomography of the dentomaxillofacial region: a systematic literature review
- PMID: 26972541
- DOI: 10.1016/j.oooo.2015.11.019
Patient movement and motion artefacts in cone beam computed tomography of the dentomaxillofacial region: a systematic literature review
Abstract
Objectives: To undertake a systematic review on the current knowledge regarding patient movement detection and patient motion artefacts related to cone beam computed tomography (CBCT) imaging of the dentomaxillofacial region.
Methods: The MEDLINE (PubMed) bibliographic database was searched for a period up to June 2015 for studies evaluating patient movement and/or motion artefacts in CBCT. The search strategy was restricted to English language publications using the following combined terms: (movement OR motion) AND (CBCT OR cone beam CT).
Results: The search strategy yielded eight publications, which qualitatively or quantitatively evaluated patient movement and/or patient motion artefacts in CBCT.
Conclusions: The literature suggests that patient movement usually presents itself in CBCT images as stripe-like and ring-like patterns, double bone contours, and overall lack of sharpness. Studies monitoring patients during CBCT examination reported a prevalence of movement in approximately 20% of the cases, and studies based on image artefact recognition to define patient movement reported prevalence as high as 41.5%. There seems to be a consensus on the fact that young patients (children and adolescents) often move during the examination.
Copyright © 2016 Elsevier Inc. All rights reserved.
Similar articles
-
Quality assurance phantoms for cone beam computed tomography: a systematic literature review.Dentomaxillofac Radiol. 2017 Mar;46(3):20160329. doi: 10.1259/dmfr.20160329. Epub 2017 Feb 17. Dentomaxillofac Radiol. 2017. PMID: 28045345 Free PMC article.
-
Computed Tomographic Artifacts in Maxillofacial Surgery.J Craniofac Surg. 2018 Jan;29(1):e78-e80. doi: 10.1097/SCS.0000000000004088. J Craniofac Surg. 2018. PMID: 29194265
-
Cone-Beam Computed Tomography for Detection of Intrabony and Furcation Defects: A Systematic Review Based on a Hierarchical Model for Diagnostic Efficacy.J Periodontol. 2016 Jun;87(6):630-44. doi: 10.1902/jop.2016.150636. Epub 2016 Feb 15. J Periodontol. 2016. PMID: 26876352
-
Impact of voxel size variation on CBCT-based diagnostic outcome in dentistry: a systematic review.J Digit Imaging. 2013 Aug;26(4):813-20. doi: 10.1007/s10278-012-9562-7. J Digit Imaging. 2013. PMID: 23254628 Free PMC article.
-
Influence of metal artefact reduction tool on the detection of vertical root fractures involving teeth with intracanal materials in cone beam computed tomography images: A systematic review and meta-analysis.Int Endod J. 2021 Oct;54(10):1769-1781. doi: 10.1111/iej.13569. Epub 2021 Jun 22. Int Endod J. 2021. PMID: 34003491
Cited by
-
Head motion during cone-beam computed tomography: Analysis of frequency and influence on image quality.Imaging Sci Dent. 2020 Sep;50(3):227-236. doi: 10.5624/isd.2020.50.3.227. Epub 2020 Sep 16. Imaging Sci Dent. 2020. PMID: 33005580 Free PMC article.
-
[A deep blur learning-based motion artifact reduction algorithm for dental cone-beam computed tomography images].Nan Fang Yi Ke Da Xue Xue Bao. 2024 Jun 20;44(6):1198-1208. doi: 10.12122/j.issn.1673-4254.2024.06.22. Nan Fang Yi Ke Da Xue Xue Bao. 2024. PMID: 38977351 Free PMC article. Chinese.
-
Correlation between magnetic resonance imaging and cone-beam computed tomography for maxillary sinus graft assessment.Imaging Sci Dent. 2020 Jun;50(2):93-98. doi: 10.5624/isd.2020.50.2.93. Epub 2020 Jun 18. Imaging Sci Dent. 2020. PMID: 32601583 Free PMC article.
-
Through the quality kaleidoscope: reflections on research in dentomaxillofacial imaging.Dentomaxillofac Radiol. 2020 Sep 1;49(6):20190484. doi: 10.1259/dmfr.20190484. Epub 2020 Feb 11. Dentomaxillofac Radiol. 2020. PMID: 31971827 Free PMC article.
-
Valid 3D surface superimposition references to assess facial changes during growth.Sci Rep. 2021 Aug 12;11(1):16456. doi: 10.1038/s41598-021-95942-3. Sci Rep. 2021. PMID: 34385558 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources