Abstract
Objectives: The aims of this study were to evaluate microbial contamination in phosphor storage plates in dental radiology services and discuss the possible origin of this contamination.
Materials and methods: The sample comprised 50 phosphor plates: 14 plates from service A, 30 from service B, and 6 in the control group, consisting of plates never used. Damp sterile swabs were rubbed on the phosphor plates, and then transferred to tests tubes containing sterile saline solution. Serial dilutions were made, and then inoculated in triplicate on Mueller Hinton agar plates and incubated at 37 °C/48 h, before counting the colony-forming units (CFU). The samples were also seeded in brain-heart infusion medium to confirm contamination by turbidity of the culture medium. All solutions, turbid and clean, were seeded in selective and non-selective media.
Results: At service A and B, 50 and 73.3 % of the phosphor plates were contaminated, respectively. This contamination was mainly due to bacteria of the genus Staphylococcus. CFU counts ranged from 26.4 to 80.0 CFU/plate.
Conclusions: Most of the phosphor plates evaluated shown to be contaminated, mainly by Staphylococcus ssp. Quantitatively, this contamination occurred at low levels, possibly arising from handling of the plates. The use of a second plastic barrier may have diminished contamination by microorganisms from the oral cavity.
Clinical relevance: There is a risk of cross-contamination by phosphor storage plates used in dental radiology services.
Keywords: Dental digital radiography; Dental radiography; Equipment contamination; Infection control.
Similar articles
-
Infection control in digital intraoral radiography: evaluation of microbiological contamination of photostimulable phosphor plates in barrier envelopes.J Can Dent Assoc. 2011;77:b93. J Can Dent Assoc. 2011. PMID: 21929948
-
An evaluation of microbiologic contamination on a phosphor plate system: is weekly gas sterilization enough?Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010 Mar;109(3):457-62. doi: 10.1016/j.tripleo.2009.09.035. Epub 2010 Jan 8. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010. PMID: 20060337
-
An evaluation of microbiologic contamination on phosphor plates in a dental school.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009 Feb;107(2):279-82. doi: 10.1016/j.tripleo.2008.05.025. Epub 2008 Jul 25. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009. PMID: 18656390
-
Digital imaging for dental caries.Dent Clin North Am. 2000 Apr;44(2):319-38, vi. Dent Clin North Am. 2000. PMID: 10740771 Review.
-
Microbial contamination of dental unit waterlines.Ann Agric Environ Med. 2008;15(2):173-9. Ann Agric Environ Med. 2008. PMID: 19061251 Review.
Cited by
-
Cinnamaldehyde is a biologically active compound for the disinfection of removable denture: blinded randomized crossover clinical study.BMC Oral Health. 2020 Aug 17;20(1):223. doi: 10.1186/s12903-020-01212-5. BMC Oral Health. 2020. PMID: 32807162 Free PMC article.
-
Intraoral digital radiography: A comprehensive report on the technical specifications of current and historical systems.Imaging Sci Dent. 2025 Mar;55(1):72-89. doi: 10.5624/isd.20240248. Epub 2025 Mar 10. Imaging Sci Dent. 2025. PMID: 40191390 Free PMC article.
-
Effect of enveloping and disinfection methods on artefact formation on enveloped PSP plate images.Oral Radiol. 2022 Oct;38(4):558-564. doi: 10.1007/s11282-022-00587-1. Epub 2022 Jan 22. Oral Radiol. 2022. PMID: 35064886 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources