Durability after aortic valve replacement with the Mitroflow versus the Perimount pericardial bioprosthesis: a single-centre experience in 2393 patients
- PMID: 26984983
- DOI: 10.1093/ejcts/ezv432
Durability after aortic valve replacement with the Mitroflow versus the Perimount pericardial bioprosthesis: a single-centre experience in 2393 patients
Abstract
Objectives: This study compares the durability and risk of reoperation in patients undergoing aortic valve replacement (AVR) with either a Mitroflow or a Carpentier-Edwards (CE) pericardial bioprosthesis. Since AVR with bioprosthetic valves has increased progressively in recent years as compared to mechanical valves, especially in patients aged 60-70 years, there has been renewed interest in the long-term durability of current pericardial bioprostheses.
Methods: We compared 440 AVR with Mitroflow valves with 1953 AVR with CE pericardial valves implanted from 1999 to 2014 with regard to reoperation, reoperation for structural valve deterioration (SVD) and all-cause mortality.
Results: Ten-year freedom from explant of any cause was higher for CE Perimount (98 ± 0.7%) than for Mitroflow (95 ± 1.4%, P < 0.01). Reasons for explant for CE Perimount were SVD (n = 2), endocarditis (n = 8) and paraprosthetic leak (n = 10). The reasons for explant for Mitroflow were SVD (n = 11), endocarditis (n = 3) SVD and pericarditis (n = 1) and paraprosthetic leak (n = 2). Ten-year freedom from explant due to SVD was higher for CE Perimount (100%) than for Mitroflow (96%) (P < 0.01). In small aortic annuli (bioprosthesis size 19-21 mm), freedom from SVD at 10 years for CE Perimount and Mitroflow was 100 versus 96%, respectively. By multivariate analysis, it was found that bioprosthesis size was not a risk factor for SVD. The choice of valve type could not be demonstrated to influence long-term survival.
Conclusions: The Mitroflow pericardial bioprosthesis provides less than optimal mid- and long-term durability compared with the CE Perimount pericardial valve, especially for small aortic diameter implants (19 and 21 mm). This study hereby confirms the existence of a real risk of valvular deterioration of the Mitroflow valve that might compromise the prognosis of the patients.
Keywords: Aortic valve replacement; Structural valve deterioration.
© The Author 2016. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.
Similar articles
-
Durability of pericardial versus porcine aortic valves.J Am Coll Cardiol. 2004 Jul 21;44(2):384-8. doi: 10.1016/j.jacc.2004.01.053. J Am Coll Cardiol. 2004. PMID: 15261935
-
Long-term results of the Mitroflow aortic pericardial bioprosthesis in over 800 patients: limited durability and mechanisms of dysfunction.Eur J Cardiothorac Surg. 2017 Aug 1;52(2):264-271. doi: 10.1093/ejcts/ezx161. Eur J Cardiothorac Surg. 2017. PMID: 28838100
-
Very long-term outcomes of the Carpentier-Edwards Perimount aortic valve in patients aged 50-65 years.Eur J Cardiothorac Surg. 2016 May;49(5):1462-8. doi: 10.1093/ejcts/ezv384. Epub 2015 Nov 2. Eur J Cardiothorac Surg. 2016. PMID: 26530269
-
Mitroflow aortic pericardial bioprosthesis--clinical performance.Eur J Cardiothorac Surg. 2009 Nov;36(5):818-24. doi: 10.1016/j.ejcts.2009.05.020. Epub 2009 Aug 22. Eur J Cardiothorac Surg. 2009. PMID: 19700338 Review.
-
Stentless aortic valve replacement: an update.Vasc Health Risk Manag. 2011;7:345-51. doi: 10.2147/VHRM.S11253. Epub 2011 Jun 2. Vasc Health Risk Manag. 2011. PMID: 21731886 Free PMC article. Review.
Cited by
-
Bioprosthetic valve fracture during valve-in-valve transcatheter aortic valve replacement.Proc (Bayl Univ Med Cent). 2020 Mar 6;33(3):317-321. doi: 10.1080/08998280.2020.1732267. eCollection 2020 Jul. Proc (Bayl Univ Med Cent). 2020. PMID: 32675945 Free PMC article.
-
Mid-term outcomes of Sapien 3 versus Perimount Magna Ease for treatment of severe aortic stenosis.J Cardiothorac Surg. 2020 Jun 29;15(1):157. doi: 10.1186/s13019-020-01203-1. J Cardiothorac Surg. 2020. PMID: 32600369 Free PMC article.
-
Long-term survival after Carpentier-Edwards Perimount aortic valve replacement in Western Denmark: a multi-centre observational study.J Cardiothorac Surg. 2021 May 14;16(1):130. doi: 10.1186/s13019-021-01506-x. J Cardiothorac Surg. 2021. PMID: 33990211 Free PMC article.
-
Treating Mitroflow dysfunction by means of an open valve-in-valve Perceval implantation.Kardiochir Torakochirurgia Pol. 2023 Mar;20(1):65-66. doi: 10.5114/kitp.2023.126105. Epub 2023 Apr 3. Kardiochir Torakochirurgia Pol. 2023. PMID: 37077454 Free PMC article. No abstract available.
-
Risk Factors for Mortality and Morbidity of Surgical Aortic Valve Replacement for Aortic Stenosis - Risk Model From a Japan Cardiovascular Surgery Database.Circ Rep. 2020 Jan 24;1(3):131-136. doi: 10.1253/circrep.CR-19-0010. Circ Rep. 2020. PMID: 33693127 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical