Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2013:2013:165782.
doi: 10.1155/2013/165782. Epub 2013 Apr 4.

Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry

Affiliations

Comparison of Respiratory Resistance Measurements Made with an Airflow Perturbation Device with Those from Impulse Oscillometry

J Pan et al. J Med Eng. 2013.

Abstract

The airflow perturbation device (APD) has been developed as a portable, easy to use, and a rapid response instrument for measuring respiratory resistance in humans. However, the APD has limited data validating it against the established techniques. This study used a mechanical system to simulate the normal range of human breathing to validate the APD with the clinically accepted impulse oscillometry (IOS) technique. The validation system consisted of a sinusoidal flow generator with ten standardized resistance configurations that were shown to represent a total range of resistances from 0.12 to 0.95 kPa·L(-1) ·s (1.2-9.7 cm H2O·L(-1) ·s). Impulse oscillometry measurements and APD measurements of the mechanical system were recorded and compared at a constant airflow of 0.15 L·s(-1). Both the IOS and APD measurments were accurate in assessing nominal resistance. In addition, a strong linear relationship was observed between APD measurements and IOS measurements (R (2) = 0.999). A second series of measurements was made on ten human volunteers with external resistors added in their respiratory flow paths. Once calibrated with the mechanical system, the APD gave respiratory resistance measurements within 5% of IOS measurements. Because of their comparability to IOS measurements, APD measurements are shown to be valid representations of respiratory resistance.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Cut-away diagram of airflow perturbation mechanism. The screened wheel rotates, allowing screened segments to briefly slow, or perturb, airflow passing through the pneumotachograph.
Figure 2
Figure 2
Schematic of the artificial respiratory model.
Figure 3
Figure 3
Pneumotachographs no. 1 and no. 2 (PT no. 1 and PT no. 2) and a Hans Rudolph standard flow resistor (HR) exhibit linear pressure and flow relationships at tested flows.
Figure 4
Figure 4
Increase in resistance in a series combination of resistors is linearly proportional to the number of resistors.
Figure 5
Figure 5
IOS R5 measurements of resistance of series combination of resistors slightly underestimated the directly measured resistance (a), whereas APD measurements of the same series combinations slightly overestimated the directly measured resistance (b). The offsets in both graphs are due to the unavoidable inherent resistance of the mechanical system without additional added resistances.
Figure 6
Figure 6
Strong linear correlation is observed between the IOS R5 measurements and the APD measurements (R 2 = 0.999) (a) as well as a strong proportional relation (R 2 = 0.985), shown in the Bland-Altman diagram (b). The dotted line represents ±1.96 standard deviation; the solid line represents the mean of the differences.
Figure 7
Figure 7
Uncalibrated APD-measured respiratory resistance at 9.8 Hz compared to IOS-measured respiratory resistance at 5 Hz for one typical subject, illustrating the high degree of linearity between the two devices. The linear regression was forced through zero. The dotted line indicates the line of identity. Points are averages of four trials. Error bars are three times the standard deviation.
Figure 8
Figure 8
Average data of calibrated APD-measured respiratory resistances from the cohort of all ten test subjects are nearly identical to IOS-measured respiratory resistances. The linear regression was forced through zero. The dotted line indicates the line of identity. Points are averages of four trials. Error bars are three times the standard deviation.
Figure 9
Figure 9
APD-measured respiratory resistance indicates higher respiratory resistance in the inhalation direction compared to the exhalation direction for one particular subject. The APD used for this measurement was uncalibrated. The dotted line indicates the line of identity. Points are averages of four trials. Error bars are three times the standard deviation. The offsets for the subject data are respiratory resistances that are located inside the respiratory system of the subject.
Figure 10
Figure 10
Airflow through the ventilator test lung system. Arrows indicate flexible plastic tubing connections.

Similar articles

Cited by

References

    1. Grinnan D. C., Truwit J. D. Clinical review: respiratory mechanics in spontaneous and assisted ventilation. Critical Care. 2005;9(5):472–484. doi: 10.1186/cc3516. - DOI - PMC - PubMed
    1. Johnson A. T. Biomechanics and Exercise Physiology: Quantitative Modeling. Boca Raton, Fla, USA: Taylor & Francis; 2007.
    1. Di Mango A. M. G. T., Lopes A. J., Jansen J. M., Melo P. L. Changes in respiratory mechanics with increasing degrees of airway obstruction in COPD: detection by forced oscillation technique. Respiratory Medicine. 2006;100(3):399–410. doi: 10.1016/j.rmed.2005.07.005. - DOI - PubMed
    1. Ducharme F. M., Davis G. M., Ducharme G. R. Pediatric reference values for respiratory resistance measured by forced oscillation. Chest. 1998;113(5):1322–1328. - PubMed
    1. Wohl M. E., Stigol L. C., Mead J. Resistance of the total respiratory system in healthy infants and infants with bronchiolitis. Pediatrics. 1969;43(4):495–509. - PubMed

LinkOut - more resources