Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2016 Oct 1;54(10):1635-43.
doi: 10.1515/cclm-2015-1142.

Verification of the harmonization of human epididymis protein 4 assays

Comparative Study

Verification of the harmonization of human epididymis protein 4 assays

Simona Ferraro et al. Clin Chem Lab Med. .

Abstract

Background: Serum human epididymis protein 4 (HE4) has gained relevance as an ovarian cancer (OC) biomarker and new automated methods have replaced the first released manual EIA by tracing results to it. We verified agreement and bias of automated methods vs. EIA as well as possible effects on patients' management.

Methods: One hundred and fifteen serum samples were measured by Abbott Architect i2000, Fujirebio Lumipulse G1200, Roche Modular E170, and Fujirebio EIA. Passing-Bablok regression was used to compare automated assays to EIA and agreement between methods was estimated by Lin's concordance correlation coefficient (CCC). The bias vs. EIA was estimated and compared to specifications derived from HE4 biological variation.

Results: Median (25th-75th percentiles) HE4 concentrations (pmol/L) were 84.5 (60.1-148.8) for EIA, 82.7 (50.3-153.9) for Abbott, 89.1 (55.2-154.9) for Roche, and 112.2 (67.8-194.2) for Fujirebio. Estimated regressions and agreements (95% confidence interval) were: Abbott=1.01(0.98-1.03) EIA-4.8(-7.5/-2.6), CCC=0.99(0.99-1.00); Roche=0.91(0.89-0.93) EIA+5.7(4.2/8.0), CCC=0.98(0.98-0.99); Fujirebio=1.20(1.17-1.24) EIA+ 2.4(-0.6/4.9), CCC=0.97(0.96-0.98). The average bias vs. EIA resulted within the desirable goal for Abbott [-3.3% (-6.1/-0.5)] and Roche [-0.2% (-3.0/2.5)]. However, while for Abbott the bias was constant and acceptable along the measurement concentration range, Roche bias increased up to -28% for HE4 values >250 pmol/L. Lumipulse showed a markedly positive bias [25.3% (21.8/28.8)].

Conclusions: Abbott and Roche assays exhibited a good comparability in the range of HE4 values around the previously recommended 140 pmol/L cut-off. For patient monitoring, however, the assay used for determining serial HE4 must not be changed as results from different systems in lower and higher concentration ranges can markedly differ.

PubMed Disclaimer

Similar articles

Cited by

Publication types

LinkOut - more resources