Multicenter Study
. 2016 Oct;88(4):571-581.
doi: 10.1002/ccd.26497. Epub 2016 Mar 31.

Closure of Secundum Atrial Septal Defects by Using the Occlutech Occluder Devices in More Than 1300 Patients: The IRFACODE Project: A Retrospective Case Series

Affiliations
Multicenter Study

Closure of Secundum Atrial Septal Defects by Using the Occlutech Occluder Devices in More Than 1300 Patients: The IRFACODE Project: A Retrospective Case Series

Nikolaus A Haas et al. Catheter Cardiovasc Interv. 2016 Oct.

Abstract

Background: The Occlutech Figulla ASD device series (OFSO) shows an improved device design for interventional ASD closure, larger follow-up series are missing.

Methods: We retrospectively reviewed the feasibility, safety, implantation properties, results, and follow-up of ASD closure using Occlutech devices over a 5 year period by establishing a multi-institutional collaborative result registry with 16 contributing centers from 11 countries (IRFACODE).

Results: In 1315 patients of all age groups (female 66.9%), successful (98%) ASD closure was performed (mean age 28.9 years, weight 52 kg, height 148.6 cm). Of the defects, 47.9% showed no or only a deficient aortic rim; in 11.9%, there was more than one defect; a septum aneurysm was present in 21.5%; and the mean implanted device size was 20.5 mm. Immediate closure was achieved in 78.6%, at discharge in 83.1%, and 96.4% and 97.3% at 6 and 12 months follow-up, respectively. During a mean follow-up of 2.7 years (in total 3597 patient years), significant complications were minimal (total = 8, <1%) with secondary device embolizations in five and AV-blocks in three patients. No erosion or death was reported.

Conclusion: ASD closure using OFSO is feasible in a large variety of patients, safe with only a minimal risk of severe side effects and especially without any aortic erosions despite a large percentage of large and complicated defects. © 2016 Wiley Periodicals, Inc.

Keywords: ASD closure; Flex; Flex II; IRFACODE project; Occlutech device.